Introduction
Page 27
1.5.13
Communications
Communication between the instrument and either a data logger, laptop or network can be
performed with the following communication connections located on the back panel (refer to Figure
5). These connections can be used for downloading data, onsite diagnostics, maintenance and
firmware upgrades.
RS232 #1
This port is designed to be used for simple RS232 communication.
RS232 #2
This port is designed to be used for simple RS232 communication or in multidrop configuration.
USB
This port can be used for instrument communications with equipment through a standard USB port.
TCP/IP (optional)
This port is best used for remote access and real-time access to instruments when a network is
available to connect with.
External I/O Port
The analog/digital port sends and receives analog/digital signals to other devices. These signals are
commonly used to activate gas calibrators or for warning alarms.
Analog Outputs
The instrument is equipped with a maximum of three analog outputs one for each measured gas. The
outputs type is menu selectable; voltage output (0 - 5 VDC) or current output (0 - 20 mA, 2 - 20 mA or
4 - 20 mA). The current output can also be configured as a voltage output of 0 - 10V, by configuring
the jumper (JP3) on the rear panel PCA.
Analog Inputs
The instrument is also equipped with three analog voltage inputs (0 - 5 VDC CAT 1) with resolution of
15 bits plus polarity.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Digital Status Inputs
The instrument is equipped with eight logic level inputs (0 - 5 VDC CAT 1) for the external control of
zero/span calibration sequences.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...