Smart Photoionization Detector Module PIDS3
User Manual
PIDS3 - User Manual (en) - 1.03.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 29 of 58
The calibration will be done with currently selected calibration method in measurement config.
Calibration data values stored sepa
rately for ‘standard’ and ‘extended’ calibration.
Connect power supply to the device at least one hour before attempting a calibration.
Preparations before Calibration / Calibration Test
To perform a calibration, you need a pressure cylinder of SPAN Gas and / or ZERO Gas, a flow reducer, a
tube, and a T-piece. The flow reducer should provide a flow of at least 500 ml/min. The T-piece should be
installed between flow reducer and sample port
–
dust filter. The open tube length on T-piece junction
should be 300
–
500 mm. (see following connection scheme for version with pump inside)
For calibration, it is recommended to use the
Calibration Kit
from accessories. (See chapter
Fehler! Verweisquelle konnte nicht gefunden werden.
Fehler! Verweisquelle konnte nicht ge
funden werden.
)