background image

 

9)   Set the value of the Low limit: Press COUNT/ENTER to increase the value, press TARE 

to decrease the value. 

10) Press MENU and then COUNT/ENTER to confirm the Low limit set point. 
11) No warning setup: no warning when there is no loading on the balance or the weight is 

lower than the lowest weight limit. 

Press Menu key, press Tare key repeatedly until it reads “InSPCT’. Press Count/Enter 
key, the display reads ‘SET HI’, press Tare key repeatedly until it reads ‘nOnrES’, press 
Count/Enter, the display reads “0” (the initialization value which indicates no warning 
when  the real  weight  is  lower than 0% of the lowest  weight).  Press  Count/Enter to 
increase the value, while press Tare key to decrease the value. Choose the disired value, 
press Menu key then the desired value glitters. Press Count/Enter key to confirm, the 
display reads “SET HI”. 

12) Enabling Inspection Mode: Press Tare key repeatedly until the display reads “EnAbLE”, 

press Count/Enter key to confirm; 

13) To Disable Inspection Mode: Press Tare key repeatedly until the display reads “dISAbL”, 

press Count/Enter key to confirm, the balance returns to weighing mode. 

14) Checking the highest and the lowest weight limits: 

a)  Press  Menu  key,  the display reads  “PrINT”,  press Tare key repeatedly until  it 

reads “InSPCT”. 

b)  Press Count/Enter key, the display reads “SET HI”. 
c)  Press Count/Enter key, the displays reads the highest weight limit. 
d)  Press Tare key, the display returns to reads “SET HI”. 
e)  Press Tare key, the display reads “SET LO”, press Count/Enter key, the display 

reads the lowest weight limit. 

Notice:

 

  To  leave  the  menu  setup  anytime  during  the  operation,  press Tare  key  until  the  display 

reads “ESC”, press Count/Enter to confirm. 

  To clear the highest and the lowest weight limits, press Tare key repeatedly until the display 

reads “CLEAr”, press Count/Enter key, then the limits value both return to zero. 

  After any modification of the highest and the lowest weight limits, the balance needs to 

enter the Inspect mode again. Follow the steps enable inspect mode. 

  The highest and the lowest weight limits will be saved in the balance. There is no need to 

to set up the limits for the next start-up of the balance. 

 

 

4. CALIBRATION 

Background

 

Among the various factors that may affect the accuracy of the balance, gravitational force is the 

most significant.  In different geographical areas on the earth, gravitational varies, and a balance 

should be calibrated in its specific location. 

The Accuris Analytical Balance can be calibrated in the lab using known reference weight. 

Содержание W3200-120

Страница 1: ...1 Precision Balance W3200 Series Operating Manual ...

Страница 2: ...2 ...

Страница 3: ... BALANCE 6 3 1 Display and Control Panel 6 3 2 Basic Weighing 7 3 3 Count Weighing 7 3 4 Percent Deviation 7 3 5 Inspect Mode 7 4 CALIBRATION 8 4 3 Manual Calibration 9 5 USER SET UP 9 6 ADDITIONAL FUNCTIONS 11 6 1 Bottom Hook Weighing 11 7 COMMUNICATION WITH A COMPUTER 11 7 1 RS 232 Connection 11 8 TROUBLE SHOOTING 12 9 CARE AND MAINTENANCE 13 8 1 Cleaning 13 ...

Страница 4: ...1 Standard weighing 2 Piece counting 3 Percentage weighing 4 Inspection mode Choice of weighing units grams carats or ounces RS 232 interface for connecting a peripheral device such as a computer or a printer Alarm to indicate malfunction Easy in lab calibration Bottom hook weighing used for density measurement 2 INSTALLATION 2 1 Unpacking CAUTION Always handle your Accuris balance with care The i...

Страница 5: ... 500g 1200g 3200g 5 000g Calibration External Readability 0 001g 0 01g Repeatability S D 0 002g 0 02g Linearity 0 002g 0 02g Stabilization time Approx 2 sec Operating temperature 10 30 Pan size 11 5cm 4 5 in 16cm 6 3in Draft Shield Yes 3 door glass enclosure No Dimensions W x H x D 20 5 x 35 x 30 5 cm 8 x 13 75 x 12 in 20 5 x 35 x 8 cm 8 x 13 75 x 3 1 in Power AC 110 to 230V 50 60 Hz 2 3 Environme...

Страница 6: ...On an unstable surface o In areas where there are corrosive vapors 2 4 Setting up your Balance CAUTION Always allow the balance to warm up in its ambient environment for 2 hours prior to use If the balance is moved from a cold environment to a warmer environment condensation may form on the internal components and external surfaces of the balance and this can cause operating and performance proble...

Страница 7: ... indicator negative indicator 3 2 Basic weighing Press the power key the balance will turn on and display 0 00g Place a weigh boat or weighing container on the weighing pan Press the TARE key and wait for the display to show 0 00g Place the sample onto the container close the draft shield doors if included on your balance The display will stabilize OK is displayed and the weight will be displayed ...

Страница 8: ... 3 Place the reference weight onto the weigh pan and close the doors 4 Press the PERCENT key 100 will be displayed 5 Remove the reference weight from the weigh pan 6 Place an object to be compared to the reference weight onto the weigh pan and close the draft shield doors 7 A percentage value of the deviation between reference and the sample will be displayed 3 5 Inspect Mode The Inspect mode is a...

Страница 9: ...eight limits a Press Menu key the display reads PrINT press Tare key repeatedly until it reads InSPCT b Press Count Enter key the display reads SET HI c Press Count Enter key the displays reads the highest weight limit d Press Tare key the display returns to reads SET HI e Press Tare key the display reads SET LO press Count Enter key the display reads the lowest weight limit Notice To leave the me...

Страница 10: ... program Press TARE key to choose the item to be set up then press COUNT ENTER to confirm To exit the setup program press the TARE key repeatedly until ESC is displayed then press COUNT ENTER to confirm Print Setups There are three print types STABLE Print print a stable reading when it is attained INSTANT Print print immediately after the print key is pressed INTERVAL Print print at predetermined...

Страница 11: ...to turn certain weighing units on or off To enable or disable certain units of measure perform the following procedure 1 Press MENU key the display reads PrInT 2 Press TARE key repeatedly until it reads unIT 3 Press COUNT ENTER key the display reads g yes which means g available for use Press COUNT ENTER to confirm To disable g as the unit press Tare key the display reads g no press COUNT ENTER ke...

Страница 12: ...eight P Print print function percent function instant print M Count counting function When a balance is connected to a computer it is suggested that immediate print be used In response to this command the balance will transmit whatever number or message appears on the balance display The string format output is shown below A B C D E F G H I J K L M A signs field usually no display as a space when ...

Страница 13: ...le length is 15 meters the load impedance of the device connected should be between 3000 and 7000 ohms with no more than 2500 pf shunt capacitance 8 TROUBLE SHOOTING Problem Possible Cause Solution No Display No power AC DC Power Adapter not functioning or incorrect specifications Plug in the AC DC adapter Replace the adapter Contact the Service Department Displayed weight value is unstable Unsuit...

Страница 14: ... sharp objects such as a pen or pencil on the keypad Do not let objects fall on the weighing pan otherwise the weighing system can be damaged Do not expose the balance to high temperatures or dust Do not disassemble the balance without proper training or instructions It is recommended to cover the balance when not in use Keep the balance clean and dry 8 1 Cleaning Unplug the AC adapter before clea...

Страница 15: ...15 ...

Страница 16: ...16 Copyright 2015 Benchmark Scientific PO Box 709 Edison NJ 08818 USA www benchmarkscientific co m ww w accuris usa com ...