7
from a reference weight.
1.
Empty the weigh pan;
2.
Press the TARE key to zero the display;
3.
“0.0000 g” will be displayed; Press MODE key till the display is %
mode.
4.
Place the reference weight onto the weigh pan, and close the doors;
5.
Press the SET key. 100% will be displayed;
6.
Remove the reference weight from the weigh pan;
7.
Place an object to be compared to the reference weight onto the
weigh pan and close the draft shield doors.
8.
A percentage value of the deviation between reference and the
sample will be displayed.
3-4 DISPLAY AND CONTROL PANEL
DISPLAY SYMBOLS
g
Gram
ct
Carat
Lb
Pound
oz
Ounce
pcs
piece counting mode
%
percentage mode
○
Indicates the weight is stable
+
positive indicator
-
negative indicator
·······
Waiting
CAL in
Calibration starting
CAL dn
Load the appropriate calibration weight
CAL ...
Calibrating
CAL up
Unload the calibration weight
CAL-no
Not calibrated properly
CAL end
Calibration complete
+
E
over the max capacity
-
E
under the readability
------
Set key valid
.
Reading the data
SAVE ---
Saving
S --- END
Saving complete