5.

 

 SETTING UP THE SMARTDOC 

Assemble the components of the SmartDoc enclosure as pictured in Section 4 of 
this manual.  Depending on the size of the gels to be imaged, or the desired detail 
of the image, the extender piece can be connected or removed.  The top, smart 
phone platform piece connects either to the top of the Extender or to the top of 
the Main Enclosure.  These parts will snap together and pull apart with gentle 
pressure. 

 

6.

 

SMARTDOC PHOTO FILTERS 

 
 
 

 
 
 

 
 
 

 
A photo filter is required to capture quality images with the SmartDoc enclosure.  
The orange photo filter, part E5001-ORANGE is included and this filter blocks the 
visible blue wavelengths coming from the illumination base or other blue light 
transilluminators.  It allows through the emission wavelengths from fluorescing 
nucleic acid stains, to allow clear pictures of gel bands. 

 

Additional filters are available separately for the SmartDoc and are recommended 
when using the enclosure on a UV Transilluminator.  These glass, band pass filters 
block UV light so the UV bulbs will not be seen in the images, and they also 
eliminate background wavelengths from the excitation source. 
 

E5001-590 

band pass photo filter, 590nm, for imaging EtBR on UV transilluminator 

E5001-535 

band pass photo filter, 535nm for imaging green stains on UV transilluminator 

 

Caution:  

Using the SmartDoc on a UV Transilluminator without an 

appropriate filter in place can cause eye or skin exposure to UV 
radiation.  The SmartDoc filters effectively block harmful UV radiation, 

and should always be properly installed when using UV illumination. 

 
 

E5001-ORANGE 

(included) 

E5001-535 

(optional) 

E5001-590 

(optional) 

Содержание Smartdoc 2.0

Страница 1: ...1 SmartDoc 2 0 Imaging Enclosure E5001 SD Instruction Manual Version 11 16 ...

Страница 2: ...ts 1 Introduction 3 2 Warnings 3 3 Unpacking 4 4 SmartDoc Enclosure Overview 4 5 Setting up the SmartDoc 5 6 SmartDoc Photo Filters 5 7 Gel Imaging with a Smart Phone 6 8 Specifications 7 9 Accessories and Consumables 7 ...

Страница 3: ...lthough blue light does not pose the same hazards as UV light the emitted blue light is very bright It is not recommended to look directly at the blue light illumination surface CAUTION When working in a laboratory environment it is always important to properly handle all reagents and chemicals When using a cell phone for capturing gel images it is very important not to contaminate the phone with ...

Страница 4: ...sure parts 1 Main enclosure hood 2 extending adapter and 3 top platform for smartphone Orange Photo Filter Any additional filters for UV ordered separately If any parts are damaged or missing please immediately contact Benchmark Scientific Customer Service at 908 769 5555 or email info benchmarkscientific com IMPORTANT Save the carton and packing materials for storing and transporting the SmartDoc...

Страница 5: ...tors It allows through the emission wavelengths from fluorescing nucleic acid stains to allow clear pictures of gel bands Additional filters are available separately for the SmartDoc and are recommended when using the enclosure on a UV Transilluminator These glass band pass filters block UV light so the UV bulbs will not be seen in the images and they also eliminate background wavelengths from the...

Страница 6: ...light from coming around the edges of the SmartDoc enclosure base 5 Place the SmartDoc imaging enclosure onto the transilluminator 6 Insert the appropriate filter into place on the top platform The side with the Accuris logo should be facing up See Section 6 for information on filters 7 Place a smartphone face down onto the top platform and align the camera lens with the filter NOTE Some protectiv...

Страница 7: ...cking Mat E4500 LD SmartGlow Loading Dye with Safe Green Stain 1 0ml E4500 LD S SmartGlow Loading Dye with Safe Green Stain Sample 20ul E4500 PS SmartGlow Safe Green Pre Stain 1 0ml E4500 PS SmartGlow Safe Green Pre Stain Sample 40ul Maximum Gel Size 15 x 15 cm Included imaging filter Orange PMMA 12mm aperture Dimensions WxDxH 23x19x22 cm Weight 0 9 kg Phone compatibility iPhone Samsung LG HTC sma...

Страница 8: ...8 Copyright 2016 Benchmark Scientific PO Box 709 Edison NJ 08818 USA Phone 908 769 5555 Email info accuris usa com www benchmarkscientific com www accuris usa com ...

Отзывы: