6.5  Maximum Power Limit

The myVolt Touch Power Supply has a maximum power output 
of 100 Watts.

Per Ohms Law:  

Power (P) = Current (I) x Voltage (V)

If the maximum power is reached during an electrophoresis run, 
the non-constant parameter will be limited to prevent exceeding 
the maximum power output

.

CAUTION:  

It is not recommended to run the power supply at or 

near the maximum wattage, as this will cause the power supply 
and gel tank to get hot.  Most standard electrophoresis buffer 
concentrations will not have such low resistance that the current 

and voltage level will come close to the maximum power limit.  

Touch Screen

Control and Display

4 Pairs of 

Electrical Outputs

Folding 

Front Feet

Recesses for 

Stacking 2 Units

AC Power Inlet

and Fuse

6.4  Automatic Cross Over

During an electrophoresis run, the power supply will attempt to 
maintain the constant voltage or amperage setting, depending on 
which parameter is set to “CONSTANT”.

If, during a run, the non-constant parameter setting (limit) is met 
due to the resistance of the load, the power supply will 
automatically “cross over” so the non-constant parameter will 
now be the constant and limiting parameter.  A red outline will 
appear around the new “constant” value.

FRONT

REAR

Rating, Serial #

Label

In CONSTANT CURRENT mode, the power supply will output 

the set mAmp current parameter, and the voltage will fluctuate 
relative to the resistance of the load (buffer concentration in the 
attached electrophoresis tanks) up to a maximum of the voltage 
setting that has been chosen.

Cooling 

Fan

Содержание E2301 Series

Страница 1: ...300V Power Supply Operating Manual This product is for research only Service and Contact Series E2301 A division of Version 3 18 Software rev 1 8 ...

Страница 2: ...s that could damage plastic Use caution when cleaning near the housing vents Do not allow liquid to enter the power supply 9 Troubleshooting Problem Causes Measures No display Power not connected Fuse blown Check cord connection and power supply Unplug and re attach power cord Check and replace fuse at power inlet Voltage or Amperage setting not reached Maximum wattage exceeded Resistance of the l...

Страница 3: ...nks that incorporate sufficient safety features to prevent electric shock Some types of shielded plugs may not fit into the output jacks and in this case an adapter 1 7 Overview of Electrical Principles Definitions Voltage The amount of potential energy between two points in a circuit Current The time rate of flow of an electric charge measured in amperes Resistance A material s opposition to the ...

Страница 4: ...trical Outputs Folding Front Feet Recesses for Stacking 2 Units AC Power Inlet and Fuse 6 4 Automatic Cross Over During an electrophoresis run the power supply will attempt to maintain the constant voltage or amperage setting depending on which parameter is set to CONSTANT If during a run the non constant parameter setting limit is met due to the resistance of the load the power supply will automa...

Страница 5: ...t voltage or constant amperage output with the non constant parameter set as a limit Press the box CONSTANT to switch between CONSTANT CURRENT and CONSTANT VOLTAGE running modes In constant voltage mode the power supply will output the set voltage and the mAmp current value will fluctuate float depending on the resistance of the load up to a maximum of the mAmp current setting that has been chosen...

Страница 6: ... water direct sunlight strong light corrosive gas high magnetic fields heaters fires and other heat sources 4 5 6 Operating guide 6 1 Touch Screen Control Up Down Value Adjust Buttons Adjustable Parameters Operation Buttons Status 4 Installation Place the power supply on a smooth level surface in an area where it will not contact liquids or chemicals The myVolt Power Supply features Dual Orientati...

Отзывы: