4-1 Manual Calibration

 

 

 

Plug in and pre-warm your balance at least 30 minutes before performing a the calibration 

procedure 

  Press the Power button to turn on the balance 

  Press the TARE key and the display should show 0.00 g 

  Press the CAL key, and the maximum capacity of the balance will be displayed (for 

example 500 g) 

  Pressing the TARE key will scroll through different calibration weight options.  Choose 

the appropriate value for the calibration weight that is available. 

  Place the calibration weight on the pan, and press the CAL key.  The display will read: 

CAL - - - during the calibration process. 

  When the calibration is finished, the display will read the value of the calibration weight, 

  Remove the calibration weight and the display should return to 0.00 g. 

  Calibration is complete. 

 

 

 

5. USER SET UPS

 

 

General Key Functions in User Setups

 

Press MENU key to enter setup program. 
Press TARE key to choose the item to be set up, then press COUNT/ENTER to confirm. 
To exit the setup program, press the TARE key repeatedly until “ESC” is displayed, then press 
COUNT/ENTER to confirm. 

 

Print Setups

 

There are three print types, 

STABLE Print: print a stable reading when it is attained. 
INSTANT Print: print immediately after the print key is pressed. 
INTERVAL Print: print at predetermined time intervals. 

The number of line feeds can also be set for label printing. 

Note 
The print function is separate from the line feed setup, i.e., set the print function first then 

re-enter the print MENU to program the number of line feeds. 

 

To set the print key function, use the following procedure: 

1

Print Menu: the balance reads “PrInT”. 

2

Press the COUNT/ENTER key, the balance reads “STAbLE”, which indicates stable 
print as the default. 

3

Press the TARE key to choose the print type and press COUNT/ENTER to confirm. 

4

Press the TARE key until it reads “InTEr”, press COUNT/ENTER, and “0 SEC” is 
displayed. Press TARE repeatedly to see the predetermined time interval and press 

Содержание W3200 Series

Страница 1: ...1 Precision Balance W3200 Series Operating Manual ...

Страница 2: ...2 ...

Страница 3: ... BALANCE 6 3 1 Display and Control Panel 6 3 2 Basic Weighing 7 3 3 Count Weighing 7 3 4 Percent Deviation 7 3 5 Inspect Mode 7 4 CALIBRATION 8 4 3 Manual Calibration 9 5 USER SET UP 9 6 ADDITIONAL FUNCTIONS 11 6 1 Bottom Hook Weighing 11 7 COMMUNICATION WITH A COMPUTER 11 7 1 RS 232 Connection 11 8 TROUBLE SHOOTING 12 9 CARE AND MAINTENANCE 13 8 1 Cleaning 13 ...

Страница 4: ...1 Standard weighing 2 Piece counting 3 Percentage weighing 4 Inspection mode Choice of weighing units grams carats or ounces RS 232 interface for connecting a peripheral device such as a computer or a printer Alarm to indicate malfunction Easy in lab calibration Bottom hook weighing used for density measurement 2 INSTALLATION 2 1 Unpacking CAUTION Always handle your Accuris balance with care The i...

Страница 5: ... 500g 1200g 3200g 5 000g Calibration External Readability 0 001g 0 01g Repeatability S D 0 002g 0 02g Linearity 0 002g 0 02g Stabilization time Approx 2 sec Operating temperature 10 30 Pan size 11 5cm 4 5 in 16cm 6 3in Draft Shield Yes 3 door glass enclosure No Dimensions W x H x D 20 5 x 35 x 30 5 cm 8 x 13 75 x 12 in 20 5 x 35 x 8 cm 8 x 13 75 x 3 1 in Power AC 110 to 230V 50 60 Hz 2 3 Environme...

Страница 6: ...On an unstable surface o In areas where there are corrosive vapors 2 4 Setting up your Balance CAUTION Always allow the balance to warm up in its ambient environment for 2 hours prior to use If the balance is moved from a cold environment to a warmer environment condensation may form on the internal components and external surfaces of the balance and this can cause operating and performance proble...

Страница 7: ... indicator negative indicator 3 2 Basic weighing Press the power key the balance will turn on and display 0 00g Place a weigh boat or weighing container on the weighing pan Press the TARE key and wait for the display to show 0 00g Place the sample onto the container close the draft shield doors if included on your balance The display will stabilize OK is displayed and the weight will be displayed ...

Страница 8: ... 3 Place the reference weight onto the weigh pan and close the doors 4 Press the PERCENT key 100 will be displayed 5 Remove the reference weight from the weigh pan 6 Place an object to be compared to the reference weight onto the weigh pan and close the draft shield doors 7 A percentage value of the deviation between reference and the sample will be displayed 3 5 Inspect Mode The Inspect mode is a...

Страница 9: ...eight limits a Press Menu key the display reads PrINT press Tare key repeatedly until it reads InSPCT b Press Count Enter key the display reads SET HI c Press Count Enter key the displays reads the highest weight limit d Press Tare key the display returns to reads SET HI e Press Tare key the display reads SET LO press Count Enter key the display reads the lowest weight limit Notice To leave the me...

Страница 10: ... program Press TARE key to choose the item to be set up then press COUNT ENTER to confirm To exit the setup program press the TARE key repeatedly until ESC is displayed then press COUNT ENTER to confirm Print Setups There are three print types STABLE Print print a stable reading when it is attained INSTANT Print print immediately after the print key is pressed INTERVAL Print print at predetermined...

Страница 11: ...to turn certain weighing units on or off To enable or disable certain units of measure perform the following procedure 1 Press MENU key the display reads PrInT 2 Press TARE key repeatedly until it reads unIT 3 Press COUNT ENTER key the display reads g yes which means g available for use Press COUNT ENTER to confirm To disable g as the unit press Tare key the display reads g no press COUNT ENTER ke...

Страница 12: ...eight P Print print function percent function instant print M Count counting function When a balance is connected to a computer it is suggested that immediate print be used In response to this command the balance will transmit whatever number or message appears on the balance display The string format output is shown below A B C D E F G H I J K L M A signs field usually no display as a space when ...

Страница 13: ...le length is 15 meters the load impedance of the device connected should be between 3000 and 7000 ohms with no more than 2500 pf shunt capacitance 8 TROUBLE SHOOTING Problem Possible Cause Solution No Display No power AC DC Power Adapter not functioning or incorrect specifications Plug in the AC DC adapter Replace the adapter Contact the Service Department Displayed weight value is unstable Unsuit...

Страница 14: ... sharp objects such as a pen or pencil on the keypad Do not let objects fall on the weighing pan otherwise the weighing system can be damaged Do not expose the balance to high temperatures or dust Do not disassemble the balance without proper training or instructions It is recommended to cover the balance when not in use Keep the balance clean and dry 8 1 Cleaning Unplug the AC adapter before clea...

Страница 15: ...15 ...

Страница 16: ...16 Copyright 2015 Benchmark Scientific ...

Отзывы: