background image

To return the balance to its initial position, remove the container and 
press  the  "TARE"  key,  which  will  return  the  balance  to  the  initial 
position of 0.00. 
 
It is advisable to use a container for weighing, as this will not damage 
the balance plate. This also protects the scale from possible shocks 
and increases its life. 
 
AUTO-ZERO key 
If  the  zero  value  changes  during  weighing,  press  the  "TARE"  key  to 
reset the display to zero.

 

 
 
CALIBRATION 
 

Place the balance on a flat surface 
Turn the scale on, when the display shows 0.00, press the "UNIT" key for 
6 seconds.   
 
The display will show the calibration value, this value may be modified by 
pressing the "TARE" key. The user must choose the weight with which to 
carry out the calibration, being able to choose between 10/ 20 or 50kg. 
Once the calibration weight has been set, press the "UNIT" key, the display 
will show the message CAL and then the selected weight value will flash 
(10/20  or  50kg).  Place  the  weight  on  the  weighing  platform  and  the 
message  "PASS"  will  appear,  then  the  calibration  procedure  will  be 
completed. 
 
If the scale is faulty and cannot be calibrated, the message "Err" will be 
displayed and it will switch off automatically. 
You can press the "ON/OFF" key to stop the calibration procedure.

 

 
 
 
 