1. Please place the electronic balance on a stable and flat table for use, do not place it on a shaking or
vibrating table, and use the four adjusting feet to keep the balance body stable. Please note that the
spirit bubble should be located in the center of the circle.
2. Please use a separate power outlet to avoid interference from other electrical appliances. The first
use should be charged for more than 12 hours.
3. When the power is on, please do not put anything on the scale plate.
4. When used, the center of gravity of the weighing object must be located in the center of the scale
plate, and the weighing object should not exceed the scope of the scale plate, to ensure its accuracy.
5. Before using the electronic balance, please heat up the engine for 15 ~ 20 minutes.
Maximum capacity
(
g
)
3000
6000
15000
30000
Minimum capacity
(
g
)
20
40
100
200
verification scale
intervale
(
g
)
1
2
5
10
Division (g)
0.1
0.2
0.5
1
Minimum sample
weight
(
g
)
2
4
10
20
Minimum single weight
(
g
)
0.15
0.3
0.75
1.5
Minimum sampling
quantity
(
PCS
)
10
10
10
10
repeatability
(
g
)
±0.2
±0.4
±1.0
±2.0
linearity
(
g
)
±0.2
±0.4
±1.0
±2.0
Display
3*LCD backlight display
Keys
20 touch keys
Accuracy
Ⅲ
Plate size
(
mm
)
295*230