- 6 -
- Where the ambient temperature falls -20°C or exceeds 60°C.
- Where the atmospheric pressure falls below 70kPa (700mbar) or exceeds 106kPa (1060mbar).
- Where the humidity is over 95% non-condensing.
- Where the unit is exposed to spray or splashing water.
- Where the unit is exposed to dust.
- Where the unit is exposed to water vapor.
- Where the unit is exposed to salty atmosphere.
- Where the unit is exposed to explosive gas.
- Where the unit is exposed to excessive shocks or vibrations.
- Where the angle of inclination of mounting surface exceeds 10 degrees.
- Where the unit is exposed to direct sunlight.
8) This equipment has been tested and found to comply with the limits for medical devices to the IEC
60601-1-2:2007. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates uses and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s)
are connected.
- Consult the manufacturer or field service technician for help.
9) Do not to touch signal input, signal output or other connectors, and the patient simultaneously.
10) a statement that MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding
EMC and needs to be installed and put into service according to the EMC information provided
in the ACCOMPANYING DOCUMENTS;
11) a statement that portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
12) Please consult a physician or a trained health professional for interpretation of measurement
results.
13) No phthalates are used for this product and its container.
14) The cuff is not made with natural rubber latex
Содержание BP250
Страница 1: ......
Страница 29: ...29 1 CAUTIONS FOR MEASUREMENT MEASUREMENT...
Страница 41: ...41 Print out the result of the measurement Using only magnetic card Magnetic Card reverse monitor...
Страница 44: ...44 Print out the result of the measurement Using only RFID card RFID card input data on the reverse monitor...
Страница 45: ...45 MAINTENANCE...
Страница 46: ...46...