19
5.2.2 Setting a manual temperature compensation value
To use manual temperature compensation, you need to enter the temperature value
of your process into the meter. This is the value at which readings will manually
temperature compensate. You can select any temperature between 0 and 100°C
(32 and 212°F). Default value is 25°C.
D
D
E
E
To select a manual temperature
compensation value:
1.
Switch the meter on. Press the MODE
key
to select measurement mode.
2.
If necessary, select ATC off as described
in section 5.2.1 on page 18. The ATC
indicator will not appear on the display.
3.
Press the CAL/MEAS key
to enter
conductivity or TDS calibration mode.
The CAL indicator will appear above the
primary display.
See figure
4.
While in conductivity or TDS calibration
mode, press the MODE key to enter tem-
perature calibration mode. The primary
display shows the current temperature
setting and the secondary display shows
the default value 25°C (77°F).
See figure
5.
Check the temperature of your sample
using an accurate thermometer.
6.
Press the
▲
or
▼
keys
to offset the
temperature to the measured value
from step 5.
See figure
7.
Press ENTER
to confirm the selected
temperature and to return to the con-
ductivity or TDS measurement mode.
The meter will now compensate
conductivity or TDS readings for the
manually set temperature.
F
F
Notes
To exit this program without confirming the manual temperature compensation value,
DO NOT press ENTER in step 7. Press CAL/MEAS instead.