1. The monitor has two memory zones (1 and 2). Each zone can store up to 60 measure-
ments.
2. To read memory values from a selected memory zone, use the User-Switching key to
select a memory zone (1 or 2) from which you want to recall values. Press the Memory
key. The first reading displayed is the average of all morning readings from the last 7 days.
3. Continue to press the Memory key to view the average of all nighttime readings from the
last 7 days.
4. Press the Memory key again to view the average of the last 3 measurements stored in
memory, and the last previously stored measurement. Every measurement comes with
an assigned memory sequence number.
5. All readings currently saved on the device can be transferred to the App using Bluetooth®
after successful completion of the pairing process. Select the desired user memory zone
and follow App instructions, the Bluetooth® transfer starts automatically. Data transfer will
be halted while performing a measurement or operating the device.
Note: The memory bank can store up to 60 readings per memory zone. When the number
of readings exceeds 60, the oldest data will be replaced with the new record.
Note: AM is defined as 4:00 AM – 11:59 AM
Note: PM is defined as 6:00 PM – 2:00 AM
1. Press the User-Switching key to select memory zone 1 or memory zone 2.
2. Press and hold the Memory key for approximately 5 seconds, then the data in the memory
zone can be erased automatically.
1. Adjust the date/time/Bluetooth® in the monitor by installing or replacing batteries or hold-
ing down the ON/OFF/START key for approximately 5 seconds under power off mode. The
display will show a blinking number showing the year.
2. Change the year by pressing the Memory key, each press will increase the number. Press
the ON/OFF/START key to confirm the entry and the screen will show a blinking number
representing the date.
3. Change the date, the hour and the minute as described in Step 2 above, using the Memory
key to change and the ON/OFF/START key to confirm the entries.
4. After adjusting the date/time, the Bluetooth® symbol ( ) and the blinking icon “
” will
be shown on the display simultaneously. Use the Memory key to choose whether auto-
matic Bluetooth® data transfer is activated (Bluetooth® symbol ( )+
) or deactivated
(Bluetooth® symbol ( )+
) and confirm with the ON/OFF/START key.
5. Press the ON/OFF/START key again, “0” will reappear as the Blood Pressure Monitor is ready
for measurement.
Pairing the Blood Pressure Monitor with your Smart-
phone
To begin using Bluetooth® for the first time, please visit the
website at http://www.accumed.ch for the initial set-up in-
structions.
1. Download and install the applicable APP onto your smart-
phone.
2. To pair this device with your smartphone, turn on the de-
vice, Bluetooth® and the App of smartphone, and follow
set-up and pairing instructions.
3. If the pairing is successful, the Bluetooth® symbol ( ) will
appear on the display and keep flashing during data transfer. The current measured value
will automatically be transferred to the App when the measurement is completed.
4. If the pairing has failed, the Bluetooth® symbol ( ) will not appear on the display and the
current measured value will not automatically be transferred to the App. In this case, the
value is saved in the selected user memory zone. Please re-pair this device with your smart-
phone and follow App instructions for Bluetooth® transfer.
Notes: 1. Unpairing your device will not delete the data from the App.
2. If you re-pair your smartphone with your blood pressure monitor, all prior reading history
stored on the phone App will be retained.
3. Bluetooth® data transfer will reduce the battery capacity.
If any abnormality will arise during use, please check the following points.
Symptoms
Check Points
Correction
No display when the
ON/OFF/START key is
pressed
Have the batteries run down?
Replace them with four new
batteries.
Have the batteries' polarities
been positioned incorrectly?
Re-insert the batteries in the
correct positions.
EE mark shown on dis-
play or the blood pres-
sure value is displayed
excessively low (high)
Is the cuff placed correctly?
Wrap the cuff properly so
that it is positioned correctly.
Did you talk or move during
measurement?
Measure again. Keep arm
steady during measurement.
Did you vigorously shake the
cuff during measurement?
Note: If the unit still does not work, return it to your dealer. Under no circumstance should
you disassemble and repair the unit by yourself.
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, hu-
midity, and direct sunlight. Avoid dropping or strongly shocking the main unit, and pro-
tect it from dust.
2. Clean the blood pressure monitor body and the cuff carefully with a slightly damp, soft
cloth. Do not press. Do not wash the cuff or use chemical cleaner on it. Never use thinner,
alcohol or petrol (gasoline) as cleaner.
3. Leaky batteries can damage the unit. Remove the batteries when the unit is not used for
a long time.
4. The unit should not be operated by children so to avoid hazardous situations.
5. If the unit is stored near freezing, allow it to acclimate at room temperature before use.
Measurement Method
Oscillometric
Measurement Range
Pressure: 30~260 mmHg; Pulse: 40~199 beats/
minute
Pressure Sensor
Semi conductor
Accuracy
Pressure: ± 3 mmHg; Pulse: ± 5% of reading
Inflation
Pump Driven
Deflation
Automatic Air Release Valve
Memory capacity
60 memories for each zone x 2 zones
Auto-shut-off
1 minute after last key operation
Permissible Operating Tempera-
ture and Humidity
10°C~40°C (50°F~104°F); 15%~85% RH;
700~1060 hPa
Permissible Transport and Stor-
age Temperature and Humidity
-10°C~60°C (14°F~140°F); 10%~90% RH;
700~1060 hPa
DC Power Source
DC 6V four AA Batteries
AC Power Source
DC 6V, ≥600mA (Plug size: outer(-) is Ø4.0, in-
ner(+) is Ø1.7)
Dimensions
146.7 (L) X 100 (W) X 74 (H) mm
Weight
285g (G.W.) (w/o Batteries)
Arm circumference
Adult: 24~40 cm (9.4”~15.7”)
Limited Users
Adult users
Type BF: Device and cuff are designed to
provide special protection against electrical
shocks.
IP Classification
IP21: Protection against harmful ingress of wa-
ter and particulate matter
*Specifications are subject to change without notice.
Recalling Values from Memory
Clearing Values from Memory
Time and Bluetooth® Adjustment
Data Transfer via Bluetooth®
Troubleshooting
Cautionary Notes
Specifications
Here are a few helpful tips to help you obtain more accurate readings:
• Blood pressure changes with every heartbeat and is in constant fluctuation
throughout the day.
• Blood pressure recording can be affected by the position of the user, his or her
physiological condition and other factors. For greatest accuracy, wait one hour
after exercising, bathing, eating, drinking beverages with alcohol or caffeine, or
smoking to measure blood pressure.
• Before measurement, it’s suggested that you sit quietly for at least 5 minutes as
measurement taken during a relaxed state will have greater accuracy. You should
not be physically tired or exhausted while taking a measurement.
• Do not take measurements if you are under stress or tension.
• Sit upright in a chair, and take 5-6 deep breaths. Avoid leaning back while the
measurement is being taken.
• Do not cross the legs while sitting and keep the feet flat on the floor during meas
-
urement.
• During measurement, do not talk or move your arm or hand muscles.
• Take your blood pressure at normal body temperature. If you are feeling cold or
hot, wait a while before taking a measurement.
• If the monitor is stored at very low temperature (near freezing), have it placed at a
warm location for at least one hour before using it.
• Wait 5 minutes before taking the next measurement.
1. Press the User-Switching key to select memory zone 1, memory zone 2 or guest
mode. After a memory zone is selected, press the ON/OFF/START key to reset the
monitor so it can start measurement in the chosen memory zone.
2. Press the ON/OFF/START key. All digits will light up, checking the display func-
tions. The checking procedure will be completed in 2 seconds.
3. After all symbols appear, the display will show a blinking “0”. The monitor is ready
to measure and will automatically inflate the cuff slowly to start measurement.
4. When the measurement is completed, the cuff will exhaust the pressure inside.
Systolic pressure, diastolic pressure and pulse will be shown simultaneously on
the LCD screen. The measurement is then automatically stored into the pre-des-
ignated memory zone.
5. In order to enhance the probability of pulse arrhythmia detection by the PARR
technology, measurement repetitions are recommended.
6. If Bluetooth® has been activated, the data is automatically transferred to the App
after successful completion of the pairing process, please see
Data Transfer via
Bluetooth®
.
This blood pressure monitor will re-inflate automatically to higher pressure if the
system detects that more pressure is needed to take a blood pressure measure-
ment.
Note: 1. This monitor automatically switches off approximately 1 minute after last
key operation.
2. To interrupt the measurement, simply press the ON/OFF/START key; the cuff will
deflate immediately.
Measurement Procedures
3. Center the tube over the middle of the arm. Press the hook and loop material together
securely. Allow room for 2 fingers to fit between the cuff and your arm. Position the
artery mark (Ø) over the main artery (on the inside of your arm) (Fig.
,
). Note: Locate
the main artery by pressing with 2 fingers approximately 2 cm above the bend of your
elbow on the inside of your left arm. Identify where the pulse can be felt the strongest.
This is your main artery.
4. Plug in the cuff connecting tube into the unit (Fig.
).
5. Lay your arm on a table (palm upward) so the cuff is at the same height as your heart.
Make sure the tube is not kinked (Fig.
).
6. This cuff is suitable for your use if the arrow falls within the solid color line as shown on
the right (Fig.
). If the arrow falls outside the solid color line, you will need a cuff with
other circumferences. Contact your local dealer for additional size cuffs.
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the C5,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
device could result.
If higher IMMUNITY TEST LEVELS than those specified in Table 9 are used, the minimum separation
distance may be lowered. Lower minimum separation distances shall be calculated using the equation
specified in 8.10.
Manufacturer’s declaration-electromagnetic immunity
The C5 is intended for use in the electromagnetic environment specified below. The customer or
the user of the C5 should assure that is used in such and environment.
Immunity
test
IEC 60601 test level
Compliance level
Electromagnetic environment-guid-
ance
Conducted
RF IEC
61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the C5 including
cables, than the recommended separa-
tion distance calculated from the equa-
tion applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz,
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC 61000-
4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
More information on EMC compliance of the device can be obtained from Accumed website: www.
accumed.ch
APP
APP
APP
APP
120
mmHg
120
mmHg
USB
PC
BP Manager
APP
APP
APP
APP
120
mmHg
120
mmHg
USB
PC
BP Manager
6. This unit is not field serviceable. You should not use any tool to open the device nor should
you attempt to adjust anything inside the device. If you have any problems, please contact
the store or the doctor from whom you purchased this unit.
7. As a common issue for all blood pressure monitors using the oscillometric measurement
function, the device may have difficulty in determining the proper blood pressure for us-
ers diagnosed with diabetes, poor circulation of blood, kidney problems, or for users suf-
fered from stroke, or for unconscious users.
8. This unit is able to detect common arrhythmia (atrial or ventricular premature beats or
atrial fibrillation). The ARR, AFib and PC icons are displayed after the measurement if Atrial
Fibrillation and Premature Contraction was detected during the measurement. If ARR,
AFib or PC icons are displayed, you are advised to wait for a while and take another meas-
urement. It is strongly recommended that you consult your physician if the ARR, AFib or
PC icons appear often.
9. While the given device is able to detect specific pulse arrhythmia, the measurement ac-
curacy of the blood pressure meter may be impaired with the occurrence of pulse ar-
rhythmia.
10. To stop operation at any time, press the ON/OFF/START key, and the air in the cuff will be
rapidly exhausted.
11. Once the inflation reaches 300 mmHg, the unit will start deflating rapidly for safety rea-
sons.
12. Please note that this unit can be a home healthcare product, but it is not intended to
serve as a substitute for the advice of a physician or medical professional.
13. Do not use this device for diagnosis or treatment of any health problem or disease. Meas-
urement results are for reference only. Consult a healthcare professional for interpreta-
tion of pressure measurements. Contact your physician if you have or suspect any medi-
cal problem. Do not change your medications without the advice of your physician or
healthcare professional.
14. Electromagnetic interference: Avoid strong electrical or electromagnetic fields in the
direct vicinity of the device (e.g.mobile telephones, microwave ovens) or less than 1,5
km from AM, FM or TV broadcast antennas. These may lead to temporary impairment of
measurement accuracy.
15. Dispose of device, batteries, components and accessories according to local regulations.
16. This monitor may not meet its performance specification if stored or used outside tem-
perature and humidity ranges specified in Specifications.
17. Please note that when inflating, the functions of the limb in question may be impaired.
18. During the blood pressure measurement, blood circulation must not be stopped for an
unnecessarily long time. If the device malfunctions, remove the cuff from the arm.
19. Avoid any mechanical restriction, compression or bending of the cuff line.
20. Do not allow sustained pressure in the cuff or frequent measurements. The resulting re-
striction of the blood flow may cause injury.
21. Ensure that the cuff is not placed on an arm in which the arteries or veins are undergoing
medical treatment, e.g. intravascular access or therapy, or an arteriovenous (AV) shunt.
22. Do not apply the cuff on the side, where a mastectomy has been performed in your
patient history.
23. Do not place the cuff over wounds as this may cause further injury.
24. Only ever use the cuffs provided with the monitor or original replacement cuffs. Other-
wise erroneous results will be recorded.
25. Batteries can be fatal if swallowed. You should therefore store the batteries and products
where they are inaccessible to small children. If a battery has been swallowed, call a doc-
tor immediately.
26. Do not use the tubing and/or AC adaptor for any other purpose than those specified, as
they can cause risk of strangulation.
27. Do not service or maintain device and cuff while in use.
28. This device should not be used adjacent to or stacked with other equipment.
29. Please do not use any other cables or accessories not approved by the manufacturer in
this manual to avoid negative influence on electromagnetic compatibility.
IN0C50000000000X
X
AM_IB_C5(1)NA17B
T_
EN_SW_v
er2138
WARNING:
The symbol on this product means that it's an electronic product and
following the European directive 2012/19/EU the electronic products have to be dis-
posed on your local recycling centre for safe treatment.
Warranty Card
This instrument is covered by a 3 year guarantee from the date of purchase. The guar-
antee is valid only on presentation of the warranty card completed or stamped by the
seller/dealer confirming date of purchase or the receipt. Batteries, cuff and accessories
are not included. Opening or altering the instrument invalidates the guarantee. The
guarantee does not cover damage, accidents or non-compliance with the instruction
manual. Please contact your local seller/dealer or www.accumed.ch.
Customer Name: ____________________________________________________
Address: ___________________________________________________________
Telephone: __________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store where purchased:
____________________________________________________________________
Rossmax Swiss GmbH,
Widnauerstrasse 1, CH-9435
Heerbrugg, Switzerland
- The
Bluetooth
® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Rossmax International Ltd. is under
license. Other trademarks and trade names are those of their respective owners.
- The blood pressure monitor uses
Bluetooth
®(
Bluetooth
® low energy technology)
- Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
- Google Play and the Google Play logo are trademarks of Google Inc.