Example
Determine an unknown piece count; weigh the preset reference sample quantity
Settings (changes in the factory settings required for this example):
Menu:
Application program: Counting (menu code
2. 1. 4
)
Reference sample quantity: 20 pcs (menu code
3. 3. 3
)
Step
Key (or instruction)
Display/Printout
1. Turn on the balance/scale
[ON/OFF]
2. Tare the balance/scale
[ZERO]
0.0
g
3. Display the reference sample
[F] >2 sec.
ref 20
(briefly)
quantity (here: 20 pcs)
4. Place the reference sample quantity
+
66.0
g
(20 pcs) on the balance/scale (here: 66 g)
5. Start the application; if the print format
[F]
+
20
pcs
is set to include data ID codes,
the following is printed:
wRef + 3.300 g
6. Weigh uncounted parts
+
174
pcs
(here: 174 pcs)
7. Display weight
[F]
+
574.2
g
8. Display quantity
[F]
+
174
pcs
9. Unload the balance/scale
0
pcs
10. Delete the reference value
[CAL] > 2 sec.
11. Repeat the procedure starting from step 6, if desired.
12
www.balances.com is your Acculab deep discount distributor