EXM-150 MICROSCOPE SERIES 

ACCU-SCOPE

®

 

73 Mall Drive, 

Commack, NY  11725  •  631-864-1000  •  www.accu-scope.com 

10

ADJUSTMENT & OPERATION

 (continued)

Replacing the Rechargeable Batteries 

1. Unplug the microscope from the electrical outlet

(if plugged in) and unplug charger from rear of microscope
(if plugged in).

2. Carefully place the microscope on the back of its arm so

the bottom of the microscope base is facing towards you.

3. To expose the battery compartment, open the trap door on

the bottom by unscrewing the lock screw 

.

4. Using a small hex wrench, remove the screw 

 on the

battery compartment and slide the cover 

 toward you to

remove.

5. Replace the three NiMH AA 1300mAh rechargeable

batteries 

.

6. Slide the cover back on and replace the screw.

7. Close the trap door and tighten the lock screw.

8. Carefully place the microscope in an upright position.

9. Plug the charger to the power receptacle on the back of

the microscope and plug into an electrical receptacle for
continued use while charging.

Allow the LED microscope to charge for 8 hours before using 
as cordless.  

Replacing the LED Bulbs

 

The life expectancy of the LED bulbs is 20,000 hours.   
The LED bulbs are covered by a 1 year manufacturer’s 
warranty.   

Should your LED bulbs need replacing, please contact an 
authorized ACCU-SCOPE service center or call ACCU-
SCOPE Inc.’s technical service department at  
631-864-1000 for an authorized service center near you.

IMPORTANT 

USE ONLY 

AA 13

00mAh NiMH (Nickel Metal 

Hydride) rechargeable batteries in your microscope. 
Using any other type of battery may damage your 
microscope

Содержание EXM-150 SERIES

Страница 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com EXM 150 MICROSCOPE SERIES...

Страница 2: ...CE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS DIAGRAMS 4 5 ADJUSMENT OPERATION CONNECTING THE POWER CORD 6 ILLUMINATION 6 SPECIMEN SLIDE 7 FOCUSING 7 DISC OR IRIS DIAPHRAGM 8 OIL OBJECTIVE 8 RECHARGING...

Страница 3: ...ponent including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More pers...

Страница 4: ...ighest quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived pre assembled and packed in a molded container Do not discard the container the molded container should be retained fo...

Страница 5: ...25 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Other Models Not Shown EXM 150 MS EXM 151 MST EXM 150 T EXM 151 MS Y Axis Adjustment Top Knob X Axis Adjustment Bottom Knob Condenser Models Fea...

Страница 6: ...pe uses a universal power supply so it may be used in any voltage range between 90 240v when used with the appropriate line cord IMPORTANT Use care when storing the power cord so that it does not bend...

Страница 7: ...a specimen slide into the slide holder with the cover glass facing up Release the slide finger so it closes and secures the slide in place Rotate the X and Y axis knobs to move the slide into positio...

Страница 8: ...cimen slide Rotate the nosepiece so the 100xR oil immersion objective is in the light path Dust or air bubbles in the oil can destroy the definition of the image If the bubbles are trapped between the...

Страница 9: ...000A line cord charger and requires approximately 8 hours for a full charge The life expectancy of the rechargeable batteries is approximately 20 000 hours or equivalent to 400 recharges before needin...

Страница 10: ...ies 6 Slide the cover back on and replace the screw 7 Close the trap door and tighten the lock screw 8 Carefully place the microscope in an upright position 9 Plug the charger to the power receptacle...

Страница 11: ...ck stop position by swinging the objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the len...

Страница 12: ...rotating the intensity control dial and or iris diaphragm Insufficient brightness Lamp intensity is too low Aperture diaphragm closed too far Condenser position too low Adjust the light intensity by...

Страница 13: ...oes not light when switched on No electrical power Lamp bulb burnt out Fuse blown out Check power cord connection Replace bulb Replace fuse Slippage of focus when using the coarse focusing knob Tensio...

Страница 14: ...orkmanship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of two years from the date of invoice to the original end user...