background image

 

3025 SERIES PHASE CONTRAST SET-UP INSTRUCTIONS 

 

1.

 

Place the microscope in front of you and plug in your light source. 

2.

 

Mount the objectives clockwise (lowest magnification to highest magnification) onto the 
microscope nosepiece. 

3.

 

Install the turret phase condenser (Fig. 2) into the condenser holder located under the 
stage and tighten into place.  

4.

 

Position the turret phase condenser so that the “BF” on the rotatable annuli selector wheel 
is  facing  you.  This  places  the  condenser  iris  diaphragm  in  the  light  path.  Rack  up  the 
condenser to the top and open the diaphragm. 

5.

 

Place a specimen slide on the stage and focus on it using the 10X objective. 

6.

 

Shift the slide to an area where the specimen is no longer visible; however, do not change 
the focus. 

7.

 

Rotate the phase turret wheel counter-clockwise until the “10” figure clicks into position. 
Note: There are two rotating wheels located at the front of the phase condenser. They  are 
used to center the annuli in the condenser or phase rings.  

8.

 

Remove one of the eyepieces of the  microscope and replace  it with the phase telescope 
(Fig. 3). Note: The upper collar of the phase telescope can be moved in or out allowing 
you to bring the phase rings into focus. 

9.

 

Looking through the phase telescope you will see a dark field of view with two different 
rings ( a & b), one bright (white) and the other black. 

10.

 

While viewing the rings through the phase telescope, adjust the two rotating wheels from 
the  turret  phase  condenser  until  the  narrower  white  ring  is  superimposed  equidistant 
within the black ring (c). 

11.

 

Remove the phase telescope from the microscope and replace the eyepiece. 

12.

 

Bring the specimen back into the field of view and refocus the image until it is sharp. The 
image will be detailed and exhibit fine structural information. 

 

NOTE:  You  must  re-center  the  phase  rings  every  time  you  change  magnification  by  using  a  
different  objective.  The  turret  wheel  must  be  rotated  so  its  number  corresponds  with  the 
objective  being  used.  To  return  to  brightfield  viewing,  rotate  the  turret  wheel  until  it  is  at  the 
“BF” position.  
 
                 

            

                                            

Отзывы: