The safest way to dispose of a used introducer
needle is to immediately place it in a sharps
disposal container to reduce the risk of accidental
pricks or injuries from loose sharps. If you cannot
find a sharps disposal container right away, you
may need to recap the needle or use a needle
clipper until you have an opportunity to dispose of
sharps in an appropriate sharps disposal container.
Never throw away loose introducer needles and
other sharps in trash cans or recycling bins, and
never flush them down the toilet.
Picture 8
Attach the luer-lock connector (K) to the pump
by screwing it firmly into the pump adapter. Fo-
llow the pump manufacturer’s User’s Manual to
prime the transfer set. Prime the transfer set un-
til insulin drips from the connector needle tip. Do
not leave air bubbles in either the cartridge or
the transfer set. Connect the transfer set to the
newly inserted head set and fill the empty space
of the cannula housing with a bolus of 1.0 units
(U100 insulin).
WARNING
You must fill the empty space of the newly
inserted head set after connecting the transfer
set. Failure to do so will result in missed insulin.
If the coupling does not click noticeably and
audibly together, use a new head set and a new
transfer set.
Содержание Ultraflex
Страница 3: ...1 en es fr Instructions for use Instrucciones de uso Instructions d utilisation ...
Страница 4: ...2 transfer set head set J I H A E F G K D C B M L Infusion set ...
Страница 14: ...12 catéter cabezal J I H A E F G K D C B M L Set de infusión ...
Страница 26: ...24 dispositif tubulaire dispositif cutané J I H A E F G K D C B M L Dispositif de perfusion ...