7.7 Setup menu standard
For all SETUP MENU STANDARD settings the Accu-Chek Spirit can save
or exit using the same simple method. For all of the features in this
section you may do the following when you are finished with a section:
if you want to
confirm
and
save the changes
– press
f
. Your insulin pump returns to the
RUN
screen.
if you want to
undo the changes
– wait for your insulin pump to return to the
RUN
screen
(timeout), or
– press
d
+
a
simultaneously (exit feature) to exit
the current function screen.
7.7.1 KeyLock function
The KeyLock function allows you to lock the four keys of your insulin
pump and serves as an additional safety measure against unintentional
activation of functions (e.g. while sleeping, during contact sports).
To make use of the KeyLock function, it must be turned on.
See the section 3.3 “Keys and key combinations” for further information.
Warning
Incorrect programming of your insulin pump may cause
inappropriate insulin delivery. Your insulin pump must
be programmed with your personal settings prior
to starting insulin pump therapy. Do not operate your
insulin pump without knowing your personal settings.
If you are not sure about your personal settings or you
are less experienced, have your doctor or healthcare
team check your personal settings.
107
STANDARD user menu
Содержание Spirit
Страница 1: ...INSULIN PUMP User Guide ...
Страница 10: ...8 ...
Страница 14: ...12 ...
Страница 26: ...24 ...
Страница 36: ...34 ...
Страница 76: ...74 ...
Страница 84: ...82 ...
Страница 122: ...120 ...
Страница 146: ...144 ...
Страница 186: ...184 ...
Страница 212: ...210 ...
Страница 218: ...216 ...
Страница 225: ......