example, mobile phones. It cannot be completely ruled out that a
device of this kind could affect your micropump system. It is
therefore recommended that you keep your micropump system
at least 30 cm away from these devices while they are turned on.
Remove the micropump from your body before you undergo
examinations or treatments such as magnetic resonance imaging
(MRI), computed tomography (CT), X-rays or ultrasound, or before
you enter areas with strong electromagnetic fields.
The micropump system is not protected against effects which
may occur when using a defibrillator on a person. If a defibrillator
was used on a person wearing the micropump system, this may
cause the micropump system to malfunction. Remove the
micropump system.
w
WARNING
j
Do not use the micropump system close to sources of
strong electromagnetic radiation and ionizing radiation.
Strong electromagnetic fields, for example, from radar or
antenna installations, high-voltage sources, X-Ray sources,
magnetic resonance and computed tomography could
interfere with the micropump system. Stop the micropump
and remove it from your body before you enter areas with
electromagnetic or ionizing radiation.
j
Do not use or store the micropump system in environmental
conditions other than those for which it is intended.
1
9
Using the Micropump System in Daily Life
man_07866364001_A_en.indb 9
28.07.2017 12:15:45
Содержание Solo
Страница 18: ...10 What is the Purpose of the Micropump System man_07866364001_A_en indb 10 28 07 2017 12 15 45 ...
Страница 38: ...30 Getting to Know the Micropump System man_07866364001_A_en indb 30 28 07 2017 12 15 51 ...
Страница 64: ...56 Setting Up the Micropump System man_07866364001_A_en indb 56 28 07 2017 12 16 04 ...