PHSA Laboratories CW Site - Point of Care
Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide
Medical Approval: Dr Benjamin Jung
Medical Approval Date: 27 Dec 2016
Version: 1.2
Implementation Date: 7/2/2019 11:55:59 PM
Folder Name: CW\Point of Care\Whole Blood Glucose
This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS
Page
10
of
12
7. Touch “Range” Field
to view the Normal,, Action, or Critical Results
Ranges.
Touch “Range , “ Out of Normal Range” , “Out of Critical Range” , “Out of
Reportable Range” Fields
to view the PHSA Normal, Critical Range, Analytical Measuring Ranges for
Neonates, Pediatrics, or Adults.
A comment is required when Patient Test is “Out of Normal Range” or if there
is an “ERROR CODE”.
Touch “Cloud” symbol in lower left corner of screen.
Pre-programmed comments appear to select.
(Up to 3 comments
can be selected)
Touch “Checkmark” in lower right to confirm selection.
Touch “Check Mark” again if satisfied with comment to exit from
screen.
Note: Once exited from Patient result/comment screen, a new comment cannot be
added later
Pre-Programmed Comments when reporting a Patient Test available to use in
Meters @ PHSA:
Procedure Error
NEO GLU over 5.5 to Chk
To Repeat PT Test
Repeated Test
Repeat with Lab Draw
Doctor Notified
To Use 2
nd
Drop
Open new Test Strips
8.
Touch “Check Mark” if satisfied with patient result to exit from
screen.
Review a Patient Test or QC Test Result
9. Touch ‘Review Results – from Main Menu Screen’.
Stored test results are displayed in a sequential list.
Touch UP or DOWN ARROW to scroll up or down in the list.
Results are grouped by date.
Touch an entry in the list to display the related details.