99
Meter and Lancing Device Cleaning and Disinfecting
8
Cleaning and Disinfecting the Lancing Device
To clean and disinfect without damaging the lancing device, follow these procedures carefully.
When to Clean and Disinfect the Lancing Device
• Clean the lancing device to remove visible dirt or other material prior to disinfecting.
• Clean and disinfect the lancing device at least once per week to remove visible dirt or other material for
safe handling.
• Clean and disinfect the lancing device before allowing anyone else to handle the lancing device, for
instance, if you have someone assisting you. Do not allow anyone else to use the lancing device.
NOTE
• Do not throw away the cap after each use. Use the approved cleaning and disinfecting product on the
cap.
• Always remove the lancet drum before cleaning or disinfecting the lancing device.
• Using cleaning and disinfecting products could result in damage to the lancing device. If you notice
any of the following signs of deterioration after cleaning and disinfecting your lancing device,
stop using your lancing device and contact the Accu‑Chek Customer Care Service Center at
1‑800‑858‑8072: residue around buttons, difficulty in priming the device, difficulty in inserting the
lancet drum.
• You might observe a slight discoloration of the lancing device after multiple cleaning and disinfecting
cycles. This does not affect the functionality of the lancing device.
What to Clean and Disinfect
The following parts of the lancing device should be cleaned and disinfected:
• The entire lancing device surface
• The cap
54988_08100918001_EN.indb 99
9/30/16 5:04 PM
Содержание Guide
Страница 2: ...54988_08100918001_EN indb 2 9 30 16 5 04 PM ...
Страница 4: ...54988_08100918001_EN indb 4 9 30 16 5 04 PM ...
Страница 20: ...18 Your New System 1 54988_08100918001_EN indb 18 9 30 16 5 04 PM ...
Страница 38: ...36 Blood Glucose Tests 3 54988_08100918001_EN indb 36 9 30 16 5 04 PM ...
Страница 64: ...62 Meter Settings 5 54988_08100918001_EN indb 62 9 30 16 5 04 PM ...
Страница 82: ...80 Review Your Data 6 54988_08100918001_EN indb 80 9 30 16 5 04 PM ...
Страница 96: ...94 Wireless Communication and Meter Pairing 7 54988_08100918001_EN indb 94 9 30 16 5 04 PM ...
Страница 104: ...102 Meter and Lancing Device Cleaning and Disinfecting 8 54988_08100918001_EN indb 102 9 30 16 5 04 PM ...
Страница 122: ...120 Technical Information 10 54988_08100918001_EN indb 120 9 30 16 5 04 PM ...
Страница 128: ...126 Notes 54988_08100918001_EN indb 126 9 30 16 5 04 PM ...
Страница 129: ...127 Notes 54988_08100918001_EN indb 127 9 30 16 5 04 PM ...
Страница 130: ...128 Notes 54988_08100918001_EN indb 128 9 30 16 5 04 PM ...
Страница 131: ...54988_08100918001_EN indb 129 9 30 16 5 04 PM ...