51
Meter Maintenance and Troubleshooting
8
(The error code and
Err
will alternate on the display.)
An electronic error occurred.
Remove the batteries, press and hold the
power/set button for at least 2 seconds, and
reinsert the batteries. Perform a blood glucose or
control test.
(The error code and
Err
will alternate on the display.)
The temperature is above or below the proper
range for the system.
Refer to the test strip package insert for system
operating conditions. Move to an area with the
appropriate conditions and repeat the blood
glucose or control test. Do not artificially heat or
cool the meter.
(The error code and
Err
will alternate on the display.)
Not enough blood or control solution was drawn
into the test strip for measurement or was
applied after the test had started.
Discard the test strip and repeat the blood
glucose or control test.
(The error code and
Err
will alternate on the display.)
Blood or control solution was applied to the test
strip before the flashing drop symbol appeared
on the display.
Discard the test strip and repeat the blood
glucose or control test.
1080_Guide Me_MANUAL_MASTER.indb 51
3/5/2018 1:28:03 PM
Содержание Guide Me
Страница 2: ...1080_Guide Me_MANUAL_MASTER indb 2 3 5 2018 1 27 39 PM ...
Страница 4: ...1080_Guide Me_MANUAL_MASTER indb 4 3 5 2018 1 27 39 PM ...
Страница 32: ...30 Control Tests 4 1080_Guide Me_MANUAL_MASTER indb 30 3 5 2018 1 27 53 PM ...
Страница 36: ...34 Review Your Data 5 1080_Guide Me_MANUAL_MASTER indb 34 3 5 2018 1 27 55 PM ...
Страница 56: ...54 Meter Maintenance and Troubleshooting 8 1080_Guide Me_MANUAL_MASTER indb 54 3 5 2018 1 28 05 PM ...
Страница 64: ...62 Technical Information 9 1080_Guide Me_MANUAL_MASTER indb 62 3 5 2018 1 28 06 PM ...
Страница 68: ...66 License and Warranty 10 1080_Guide Me_MANUAL_MASTER indb 66 3 5 2018 1 28 06 PM ...
Страница 70: ...68 Index 1080_Guide Me_MANUAL_MASTER indb 68 3 5 2018 1 28 06 PM ...
Страница 71: ...1080_Guide Me_MANUAL_MASTER indb 69 3 5 2018 1 28 06 PM ...