Beep tone

Press and hold down

(for more than 3 seconds)

Time &

Date

Time-and-

date format

Hours

Minutes

Time

Date

Year

Month

Day

Press briefly

(turn off)

On

(On1)

Off

(OFF)

Acoustic Mode

(On2)

Press briefly

(turn off)

Target range

Off

On

Value

Lower limit

(LL)

Off

On

Value

Upper limit

(LH)

Press briefly

(turn off)

Alarm clock function

Press briefly

(turn off)

Press briefly

(turn off)

Turning on the meter

and starting Quick

set-up

Off

On

Hours

Minutes

AL1

AL2

AL3

AL4


Top and sides

 

Þ

 button: turn on or off, start a set-up to 

change settings, save entries

  Liquid crystal display (LCD)

 

<

 and 

>

 buttons: change settings (decre-

ment and increment value, respectively), 
retrieve results, request averages

  Test strip guide – this is where you insert the 

test strip

  Measurement optics cover

  Test strip ejector

  Code chip

  Infrared port for downloading results to a PC, 

for instance

Bottom

  Code chip

  Code chip slot

  Type plate

  Battery compartment

  Battery (type CR 2032)

  1

  Set-up

  2

  When an average is displayed: number of days 

covered

  3

  LH or LL: target range setting (set-up)

  4

  Alarm clock function turned on.  

Flashing: optical signal during alarm

  5

  Test or averaging in progress

  6

  Unit of measurement for BG result

  7

  Flag indicating special result

  8

  Flashing: cue to allow test strip to draw up 

blood

  9

  Flashing: cue to insert test strip

10

  Test strips are past their expiry date

11

  Code number display. 

Flashing: code chip is missing

12

  Target range setting (set-up). 

Flashing: result is outside the target range

13

  Flag indicating quality control

14

  Flashing: temperature is/was outside the 

permitted range

15

  Beep tone or Acoustic Mode is turned on

16

  Flashing: battery is running low

17

  When an average is displayed: number of 

results included

18

  Memory, averages or results download

19

  Follows time in 12-hour format

20

 

Display messages

  Code chip is missing.  

In memory: no results in memory, averaging 
not possible or result erased

  Alarm setting (set-up)

  Flashing: latest result being erased

  End of set-up, self-test following test or 

download complete

 

In Quick set-up: group selected

 

Result >600 mg/dL (33.3 mmol/L)

  Result <10 mg/dL (0.6 mmol/L)

  Results download

Error messages

E-1

   Test strip inserted incorrectly, test strip 

bent/moved, strong electromagnetic 
field

E-2

   Code chip from another system, meter 

cannot read code chip, code chip defec-
tive, code chip removed during test

E-4

   Measurement optics dirty, test strip 

drew up blood too soon, test strip bent/
moved, used strip inserted, test strip 
past its expiry date, new test strip inser-
ted too soon after last test (while End 
was showing)

E-6

   Test strip drew up too little blood, test 

strip drew up blood too soon, test strip 
bent/moved, battery empty, meter error, 
strong electromagnetic field

EEE

   Meter error

ttt

   Temperature too low or too high to test.

Quick set-up: Changing settings

Þ

  press and held down for more than 3 seconds at 

anytime = save, exit set-up and turn off

< >

 change with 

<

 or 

>

press briefly = save or scroll

17

18

16

15

14

13

7

6

5

4

20

3

2

1

11 10 9

8

19

12