Performing a Control Test
To perform a control test, you need the meter, a test strip and control solution
Level 1 and/or Level 2. The control level is printed on the bottle label.
1. Insert the test strip into the meter in
the direction of the arrows. This turns
the meter on automatically.
2. Make sure the code number on the
display matches the code number on
the test strip container. If you miss
seeing the code number, remove the
test strip and reinsert it into the meter.
55