- 30 -
OMNISTIM
®
FX
2
PRO USER MANUAL
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
C.
LE TRIPHASIC PENS
This program is used to re-educate lower extremity agonist - antagonist muscle pairs where fine motor
control and positioning is required. It is called triphasic because the agonist fires first, followed by the
antagonist followed by a small burst of agonist muscle to slow down the antagonist and position the joint.
Patterned Electrical Neuromuscular Stimulation (PENS) is a precise reproduction of EMG firing patterns of muscle
agonist and antagonist pairs. The stimulation helps to re-establish proper firing sequence in the spinal and
supraspinal central pattern generators, re-educating neuromuscular function. PENS helps re-boot the sense position
system and neuromuscular junction to improve recruitment and muscle function. It should be used prior to Therex
to improve recruitment and motor performance. It’s also used to decrease spasticity and muscle tone.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
Tri Phasic
D.
TENS-LVPC MOTOR
This program motor level stimulation activates recovery of function in post ischemic stroke patients.
Target tissue is superficial. Bipolar placement over local and distal acu points. The intensity should be set
to elicit a moderate muscle twitch.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…300 (μs)
200 (μs)
Y
3
PULSE RATE
0.0…250 (Hz)
4.0 (Hz)
N
PHASE DURATION MOD
20 (%)
N
MODULATION
OFF
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
A/B SIM.