2.2. General tab
2.2.1. General tab functionalities
Tab. 5–2: General tab function Keys
2.2.2. Automatic numbering
When the instrument has not received an order, the instrument then associates an automatic
Sample ID number that is incremented for each new cycle. This automatic Sample ID number
is identified in the ID field as «AUTO_SIDn» where «n» is the incremented number that is
entered in this box.
Select the «Edit» key and then select the «AUTO_SID» box figure (see
Now type in a beginning number to start the daily sequencing then select the «OK» key to
confirm your entry.
In order to enter the initial daily auto-numbering start sequence, the Worklist from
the previous day must be erased (See Section 2,
2.2.3. RUO parameters
The RUO parameters (Research Use Only) are as indicated: PCT, PDW, ALY, LIC.
If the «Visible» box on the soft parameter screen has a Check mark (see
the instrument software will trigger the following message when results are displayed on the
screen: «
PCT, PDW, ALY and LIC are for research use only
». his message will also be printed out
and/or transmitted to a host computer.
Select the «Edit» key, and then place a Check mark in the «Visible» or «Invisible» box.
Now select the «OK» key to confirm your choice.
Key/Heading
Name
Function
Auto Numbering
Auto Numbering
Defines the initial value of the autonumbered Sample
ID (see
RUO parameters
RUO parameters
Enables/disables RUO parameters for printing and
sending operations (see
Logs
Logs
Enables/disables automatic prompts for logs
(see
Default test
Default test
Test performed when no order is associated to a sample
tube.
Barcode Option
Barcode Option
Depends on laboratory operation mode
(all sample identified with barcode or not;
see
Alarms
Alarms
QC or XB alarms options Section 4,
Are the «QC failed» and «XB» blocking alarms ?
Number of consecu-
tive analyses with
Stopping conditions
Defines the number of consecutive instrument mishaps
or alarms for stopping analysis operations (from 1 to 5
times).
Operator can choose the number of consecutive trig-
gers of each alarm. These are selected by default. They
can be disabled.
Содержание Pentra 80 RAB108EA
Страница 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Страница 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Страница 34: ...Pentra 80 1 22 Pentra 80 User Manual RAB108EA Fig 1 18 Order management Patient Sample Service Order LIS P80...
Страница 40: ...Pentra 80 1 28 Pentra 80 User Manual RAB108EA 6 Printer Use the printer supplied or approved by ABX DIAGNOSTICS...
Страница 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Страница 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Страница 176: ...Pentra 80 5 24 Pentra 80 User Manual RAB108EA 5 4 2 Printer properties Fig 5 22 Printer properties window...
Страница 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Страница 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Страница 256: ...Pentra 80 7 42 Pentra 80 User Manual RAB108EA 7 Hydraulic Diagram See Hydraulic diagram on next page...
Страница 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Страница 284: ......
Страница 286: ......
Страница 288: ......
Страница 290: ......
Страница 292: ......
Страница 294: ......
Страница 296: ......
Страница 297: ...Index Glossary Pentra 80 User Manual RAB108EA 9 1 Contents 1 Glossary 9 2 2 Index 9 4 Section9 Glossary...
Страница 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...