Absolute Process Instruments Cecomp Test+ CTP3B Скачать руководство пользователя страница 4

Instructions CTP3B

Sensor 

Suggested units for calibration

3 PSI 

6.920 FTH2O

5 PSI 

5.000 PSI

15 PSI 

775.7 MMHG or TORR

30 PSI 

61.08 INHG

50 PSI 

50.00 PSI

60 PSI 

60.00 PSI

100 PSI 

7.031 KG/CM2

200 PSI 

407.2 INHG

300 PSI 

610.8 INHG

500 PSI 

3447 KPA

1000 PSI 

6895 KPA

2000 PSI 

4614 FTH2O

3000 PSI 

6920 FTH2O

5000 PSI 

5000 PSI

The display will then indicate the currently applied pressure in the 
engineering units selected for calibration.

s

 

and

 

t

 Button Operation

Each time one of the 

s

 or 

t

 buttons is pressed and released quickly, 

a small change is made to the digitized pressure signal. It may take 
more than one button press to make a change on the display. 
To make larger changes, press and hold the appropriate button. 
After about one second, the display will begin to change continu-
ously. Release the button to stop. Then make fine adjustments by 
pressing and quickly releasing the buttons as previously described.

Gauge Reference Pressure Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 
character display will alternate between ZERO and CAL. Adjust for a 
display indication of zero using the 

s

 and 

t

 buttons.

Note: At this point you may re-zero the gauge without doing any 
other calibration. Press and hold the Power button until the display 
indicates - - - - then release the button to store the new zero in 
non-volatile memory and restart the gauge.
Apply full-scale pressure. The character display will alternate 
bSPAN and CAL. Adjust for a display indication of full-
scale pressure using the 

s

 and 

t

 buttons. 

Apply 50% full-scale pressure. The character display will alternate 
between  +MID and CAL. Adjust for a display indication equal to 
50% of full-scale pressure using the 

s

 and 

t

 buttons. 

Press and hold the Power button until the display indicates - - - - 
then release the button to store the calibration parameters in non-
volatile memory and restart the gauge.
Verify readings at 0%, 25%, 50%, 75% and 100% of full scale.

Gauge Reference Vacuum Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 
character display will alternate between ZERO and CAL. Adjust for a 
display indication of zero using the 

s

 and 

t

 buttons.

Note: At this point you may re-zero the gauge without doing any 
other calibration. Press and hold the Power button until the display 
indicates - - - - then release the button to store the new zero in 
non-volatile memory and restart the gauge.
Apply full-scale vacuum. The character display will alternate 
bSPAN and CAL. Adjust for a display indication of full-
scale vacuum using the 

s

 and 

t

 buttons. 

Apply 50% full-scale vacuum. The character display will alternate 
between  +MID and CAL. Adjust for a display indication equal to 
50% of full-scale vacuum using the 

s

 and 

t

 buttons. 

Press and hold the Power button until the display indicates - - - - 
then release the button to store the calibration parameters in non-
volatile memory and restart the gauge.
Verify the readings at 0%, 25%, 50%, 75% and 100% of full scale.

Absolute Reference Gauges

Apply full vacuum to the gauge. The character display will alternate 
between 

ZERO

 and 

CAL

. Press the 

s

 and 

t

 buttons to obtain a 

display indication of zero.
Note: At this point you may re-zero the gauge without doing any 
other calibration. Press and hold the Power button until the display 
indicates - - - - then release the button to store the new zero in 
non-volatile memory and restart the gauge.
Apply full-scale pressure. The character display will alternate 
bSPAN and CAL. Press the 

s

 and 

t

 buttons to obtain a 

display indication equal to full-scale pressure.
Apply 50% of full-scale pressure. The lower display will alternate 
bMID and CAL. Press the 

s

 and 

t

 buttons to obtain an 

indication equal to 50% of full-scale pressure.
Press and hold the Power button until the display indicates - - - - 
then release the button to store the calibration parameters in non-
volatile memory and restart the gauge.
Verify the readings at 0%, 25%, 50%, 75% and 100% of full scale.

Battery Replacement 

A low battery indication will be shown in the upper left-
hand corner of the display when the battery voltage falls 
sufficiently. The batteries should be replaced soon after the 
indicator comes on or unreliable readings may result.
1.  Remove the 6 Phillips screws on the back of the unit. 
2.  Lift up the battery holder if necessary.
3.  Remove batteries by lifting up the positive end of the battery 

(opposite the spring) taking care not to bend the battery holder 
spring.

4.  Discard old batteries properly, do not discard into fire, sources 

of extreme heat, or in any hazardous manner.

5.  Always replace both batteries at the same time with high quality 

alkaline batteries.

6.  Install batteries with correct orientation. Incorrect polarity will 

damage the gauge. The negative (flat) end of each battery 
should be inserted first facing the battery holder spring.

7.  Replace battery holder and back cover, including the rubber 

gasket and reinstall the six screws.

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Calibration—Setup and Preparation

Gauges are calibrated at the factory using equipment traceable 
to NIST. There is no need to calibrate the gauge before putting 
it into service. Calibration should only be performed by qualified 
individuals using appropriate calibration standards and procedures. 
Calibration intervals depend on your quality control program 
requirements, although many customers calibrate annually. 
The calibration system must be able to generate and measure 
pressure/vacuum over the full range of the gauge and should be 
at least four times more accurate than the gauge being calibrated.
A vacuum pump able to produce a vacuum of 100 microns (0.1 torr 
or 100 millitorr) or lower is required for vacuum and absolute gauges. 
Allow the gauge to acclimate to ambient temperature for 20 minutes. 
Install fresh batteries.

Entering Calibration Mode

With the gauge off, press and hold the 

t

 button, then press the 

Power button. 
Release all buttons when the display indicates CAL.
The display begins by indicating the full-scale positive pressure 
rating of the gauge in the engineering units as configured by the 
factory, and then shows all display.
Before the gauge enters the Calibration Mode, the display initially 
indicates _ _ _ _ with the first underscore blinking, and with CALPC 
(calibration passcode) on the lower display. 
Enter the 3510 passcode as described in the Configuration 
Passcode section.

Calibration Mode

The gauge remains in the Calibration Mode until restarted manually 
or power is removed. Features not related to calibration are dis-
abled and compound range models are set for the same engineer-
ing units for pressure and for vacuum.
The calibration may be performed in any of the available engineer-
ing units as well as percent (PCT). 
For greatest accuracy, use the 

s

 and 

t

 buttons to select engi-

neering units for calibration with highest number of display counts. 
Press and release the Power button when the appropriate engineer-
ing units are displayed. Suggested units are above right.

Compound and Bipolar Gauges

In addition to the steps described above for pressure gauges, apply 
full-scale vacuum. The character display will alternate between 
-SPAN and CAL. Adjust for a display indication of actual applied 
vacuum using the 

s

 and 

t

 buttons.

For bipolar and –30.15.00psig compound range models 
only, apply 50% full-scale vacuum. The character display will alter-
nate between -MID and CAL. Adjust for a display indication equal to 
50% of full-scale vacuum using the 

s

 and 

t

 buttons. 

User-Defined Passcode Configuration 

The factory default passcode 3510 may be changed to a different 
value for configuration and/or calibration.
With the unit off, press and hold the 

s

 button to view and/or 

change the user configuration passcode. Then press the Power 
button. Release all buttons when the display indicates CFG.

Calibration Passcode

With the unit off, press and hold the 

t

 button to view and/or 

change the user calibration passcode. Then press the Power button. 
Release all buttons when the display indicates CAL.

Change Passcode Mode

Before the unit enters the view or change passcode mode, the 
display initially indicates _ _ _ _ with the first underscore blinking, 
and with CFGPC or CALPC on the character display. 
Note: The unit will automatically revert to normal operation if no 
buttons are operated for approximately 15 seconds. To cancel and 
return to normal operation, press and release the Power button.
Enter access code 1220:
Use the 

s

 and 

t

 buttons to set the left-most digit to 1.

Press and release the Power button to index to the next position. 
The 1 will remain, and the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the Power button to index to the next position. 1 
2 will remain, and the third position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the Power button to index to the next position. 1 
2 2 will remain, and the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the Power button to proceed. 
Note: If an incorrect access code was entered, the gauge will return 
to the start of the access code entry sequence.
Once the access code has been entered correctly, the display will 
indicate the existing user-defined passcode with CFGPC or CALPC 
on the character segments.
Press the 

s

 or 

t

 button to select the first character of the new 

passcode.
When the correct first character is being displayed, press and 
release the Power button to proceed to the next passcode character.
Repeat above until the entire passcode is complete.
To exit the User Defined Passcode change mode, press and hold 
the Power button. 
Release the button when the display indicates - - - -  to restart 
the gauge.

Disposal

Disposal of Electrical & Electronic Equipment 
(Applicable throughout the European Union and 
other European countries with separate collec-
tions programs). 
This symbol, indicates that this product should 
not be treated as household waste when you 
wish to dispose of it. 
Instead, it should be sent to an applicable collec-
tion point for the recycling of electrical and electronic equipment. 
The recycling of materials will help to protect and conserve natural 
resources and prevent potential negative effects on human health 
and the environment, which could otherwise arise from inappropri-
ate waste handling.
Please contact your local waste disposal authority for information 
about recycling this product and the location of your nearest des-
ignated collection point. 
You may also return this product to via pre-paid shipping to 
Absolute Process Instruments or your supplier for proper disposal.

Содержание Cecomp Test+ CTP3B

Страница 1: ...onnector to install or remove gauge without tools 304 stainless steel urethane seal SCR14SS Filter screen fitting keeps debris out of gauge sensor For food vacuum packaging applica tions 303SS body 100 micron 304SS screen Accessories order separately DPG OK2 DPG OK3 DPG OK6 Pelican brand high visibility orange heavy duty waterproof cases Models available for storing 2 3 or 6 gauges Vacuum Range Re...

Страница 2: ...nce models only Absolute reference gauges do not use the zero feature since they read atmospheric pressure under normal conditions Be sure the gauge is in the normal operating mode The gauge port must be exposed to normal atmospheric pressure with no pres sure or vacuum applied cecomp com 1220 American Way Libertyville IL 60048 800 942 0315 CECOMP Div of ABSOLUTE PROCESS INSTRUMENTS o o o o psig 0...

Страница 3: ...ear Memory _ _ _ 3 CFGPC or _ _ 5 3 CFGPC Zero Clear Memory _ _ _ 3 CFGPC _ I 5 3 CFGPC Zero Clear Memory 0 I 5 3 CFGPC t s t s t s Zero Clear Memory USER Zero Clear Memory FCTRY or t s Zero Clear Memory USER t s t s MAN r l c Auto man Zero Clear Memory Zero Clear Memory t s or t s Engineering Unit Selection The upper display will be blank with the engineering units in the lower display Use the s ...

Страница 4: ... be performed by qualified individuals using appropriate calibration standards and procedures Calibration intervals depend on your quality control program requirements although many customers calibrate annually The calibration system must be able to generate and measure pressure vacuum over the full range of the gauge and should be at least four times more accurate than the gauge being calibrated ...

Отзывы: