Absolute Process Instruments Cecomp Test+ CTP3B Скачать руководство пользователя страница 1

Cecomp

®

 

Test +

 Battery Powered Digital Pressure Gauges, User Configurable 

CTP3B

Ranges and Resolution

See table below for ranges 
Default engineering unit: psi
User configurable for other engineering units 
Resolution is fixed for each engineering unit

Accuracy

Accuracy includes linearity, hysteresis, repeatability 
Sensor hysteresis:  ±0.015% FS, included in accuracy 
Sensor repeatability:  ±0.01% FS, included in accuracy

Pressure and absolute ranges

±0.1% full scale ±1 least significant digit

Vacuum and vacuum/pressure ranges

±0.25% full scale ±1 least significant digit

Display

3 readings per second nominal display update rate 
4 digit LCD, 0.5"H and 5 character 0.25"H alphanumeric 
White LED backlight active with button keypress

Batteries, Battery Life, Low Battery Indication

2 AA alkaline included

 

Approx. 150-1500 hours depending on backlight usage 
Low battery symbol on display

Controls and Functions

Three button keypad 
Zero/clear/up 
On/off 
Memory/down
Passcode protected calibration, engineering unit selection, auto 
shutoff time, min/max setup

Min/Max Memory

Minimum and/or maximum readings stored in memory 
Readings cleared or stored at shutoff 
User configurable

Calibration

Zero button for gauge reference ranges 
Non-interactive zero, span, and linearity, ±10% of range

Auto Shutoff

Factory default: 5 minutes  
User selectable 1 minute to 8 hours or front button on/off

Weight

Gauge: 

9 ounces (approximately) 

Shipping: 

1 pound (approximately)

Housing Materials

Epoxy powder coated aluminum case with case stiffener, 
ABS/ polycarbonate bezel, coated circuit boards for humidity 
resistance, front and rear rubber gaskets, polycarbonate label 
NEMA 2 
Includes rubber boot

Connection and Material

1/4" NPT male fitting 
All wetted parts are 316L stainless steel

Overpressure, Burst, Vacuum Service

3000 psig sensor:  5000 psig overpressure 
5000 psig sensor:  7500 psig overpressure 
All others: 

2 X pressure range overpressure 

Burst pressure:   4 X sensor pressure rating, or 10,000 psi,  
 

whichever is less 

Vacuum service:  15 psia, ±15 psig, 15 psig, 30 psia,  
 

100 psig, 100 psia, 200 psig sensors

Environmental Temperatures

Storage: 

–40 to 203°F  (–40 to 95°C) 

Operating: 

  –4 to 185°F  (–20 to 85°C) 

Compensated range:   32 to 158°F      (0 to 70°C)

Dimensions

3.67"W x 3.19"H x 2"D with boot, not including fitting 

© 10-19

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

cecomp.com/plus

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

Made in USA

Pb

Lead 

Free

 

O

±0.1% Test Gauge Accuracy in Most Ranges

 

O

Ruggedized Design

 

O

Red Rubber Boot Included

 

O

316L Stainless Steel Wetted Parts

 

O

Selectable Units and Auto Shutoff Times

 

O

White LED Display Backlight

 

O

Min/Max Memory

CON14SS

Quick connector to install or remove gauge 
without tools. 304 stainless steel, urethane 
seal.

SCR14SS

Filter screen fitting keeps debris out of gauge 
sensor. For food vacuum packaging applica-
tions. 303SS body, 100 micron 304SS screen. 

Accessories—order separately

DPG-OK2, DPG-OK3, DPG-OK6

Pelican

®

 brand high visibility orange heavy 

duty waterproof cases. Models available for 
storing 2, 3, or 6 gauges.

Vacuum

Range/Resolution

Accuracy

Selectable Units

CTP3B15PSIVAC

 

0 to 14.70 psig vacuum  

±0.25% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2 cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

Vacuum/Pressure

Range/Resolution

Accuracy

Selectable Units

CTP3B±15PSIG

  –14.7 to 15.0 psig

 

±0.25% FS ±1 LSD

psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B15V100PSIG

  –14.7 to 100.0 psig

 

±0.25% FS ±1 LSD

psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, MPa, kPa

CTP3B15V200PSIG

  –14.7 to 200.0 psig

 

±0.25% FS ±1 LSD

psi, inHg/psi, inHg, torr inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa

Pressure

Range/Resolution

Accuracy

Selectable Units

CTP3B3PSIG

 

0 to 3.000 psig

 

±0.25% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa

CTP3B5PSIG

 

0 to 5.000 psig

 

±0.25% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa

CTP3B15PSIG

 

0 to 15.00 psig

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B30PSIG

 

0 to 30.00 psig

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B60PSIG

 

0 to 60.00 psig

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B100PSIG

 

0 to 100.0 psig

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B200PSIG

 

0 to 200.0 psig

 

±0.1% FS ±1 LSD

psi, inHg, inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa

CTP3B300PSIG

 

0 to 300.0 psig

 

±0.1% FS ±1 LSD

psi, inHg, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa

CTP3B500PSIG

 

0 to 500.0 psig

 

±0.1% FS ±1 LSD

psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa

CTP3B1000PSIG

 

0 to 1000 psig

 

±0.1% FS ±1 LSD

psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa

CTP3B2000PSIG

 

0 to 2000 psig

 

±0.1% FS ±1 LSD

psi, inHg, ftH2O, kg/cm2, atm, bar, MPa

CTP3B3000PSIG

 

0 to 3000 psig

 

±0.1% FS ±1 LSD

psi, inHg, ftH2O, kg/cm2, atm, bar, Mpa

CTP3B5000PSIG

 

0 to 5000 psig

 

±0.1% FS ±1 LSD

psi, kg/cm2, atm, bar, Mpa

Absolute Reference

Range/Resolution

Accuracy

Selectable Units

CTP3B15PSIA

  15.00 to 0 psi absolute

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B30PSIA

  30.00 to 0 psi absolute

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

CTP3B100PSIA

  100.0 to 0 psi absolute

 

±0.1% FS ±1 LSD

psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa

Calibration Documentation Options—add to end of model number

CD

Calibration data; 5 test points and date 

NC

NIST traceability documentation, 5 points and date

Содержание Cecomp Test+ CTP3B

Страница 1: ...onnector to install or remove gauge without tools 304 stainless steel urethane seal SCR14SS Filter screen fitting keeps debris out of gauge sensor For food vacuum packaging applica tions 303SS body 100 micron 304SS screen Accessories order separately DPG OK2 DPG OK3 DPG OK6 Pelican brand high visibility orange heavy duty waterproof cases Models available for storing 2 3 or 6 gauges Vacuum Range Re...

Страница 2: ...nce models only Absolute reference gauges do not use the zero feature since they read atmospheric pressure under normal conditions Be sure the gauge is in the normal operating mode The gauge port must be exposed to normal atmospheric pressure with no pres sure or vacuum applied cecomp com 1220 American Way Libertyville IL 60048 800 942 0315 CECOMP Div of ABSOLUTE PROCESS INSTRUMENTS o o o o psig 0...

Страница 3: ...ear Memory _ _ _ 3 CFGPC or _ _ 5 3 CFGPC Zero Clear Memory _ _ _ 3 CFGPC _ I 5 3 CFGPC Zero Clear Memory 0 I 5 3 CFGPC t s t s t s Zero Clear Memory USER Zero Clear Memory FCTRY or t s Zero Clear Memory USER t s t s MAN r l c Auto man Zero Clear Memory Zero Clear Memory t s or t s Engineering Unit Selection The upper display will be blank with the engineering units in the lower display Use the s ...

Страница 4: ... be performed by qualified individuals using appropriate calibration standards and procedures Calibration intervals depend on your quality control program requirements although many customers calibrate annually The calibration system must be able to generate and measure pressure vacuum over the full range of the gauge and should be at least four times more accurate than the gauge being calibrated ...