Negative Midpoint Non-Linearity Calibration (bipolar rang-
es only)

Apply 50% full-scale negative pressure.
The lower display will alternate between CAL and MID.
Use the 

s

 and 

t

 buttons to adjust the upper display to indi-

cate the applied pressure value.

Exit Calibration Mode

To store the calibration parameters and exit calibration mode, 
press and hold the SEL button until the display indicates  
– – – – .

Configuration Pass Code

From the normal operating mode, press and hold the TEST and 
the 

s

 buttons. Then press the SEL button. Release all buttons 

when the display indicates CFG.

Calibration Pass Code

From the normal operating mode, press and hold the TEST and 
the 

t

 buttons. Then press the SEL button. Release all buttons 

when the display indicates CAL.

Test and Set Point Adjust Pass Code (optional)

From the normal operating mode, press and hold the TEST 
button.
Then press the SEL button.
Release all buttons when the display indicates  – – – – .
Before the unit enters the view or change pass code mode, 
the display initially indicates  _ _ _ _  with the first underscore 
blinking, and with CFGPC, CALPC, or TSTPC on the lower 
display.
Note: during pass code entry, the LEDs will extinguish and 
the gauge will not respond to changes in applied pressure. 
The output relays will maintain their prior states and the 
retransmission output will maintain its prior value. The gauge 
will automatically revert to normal operation if no buttons are 
operated for approximately 15 seconds.

Enter Factory Pass Code 1220

Use the 

s

 and 

t

 buttons to set the left-most digit to 1.

Press and release the SEL button to index to the next position. 
The 1 will remain, and the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the SEL button to index to the next position. 1 
2 will remain, and the third position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the SEL button to index to the next position. 1 
2 2 will remain, and the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the SEL button to proceed. Note: If an incor-
rect pass code was entered, the gauge will exit to the normal 
operating mode.
Once the correct password has been entered, the display will 
indicate the existing pass code with CFGPC, CALPC, or TSTPC 
on the character segments. 
Note: while in the pass code change mode, the LEDs will 
extinguish and the unit will not respond to changes in applied 
pressure and the output relays will be de-energized.
Operate the 

s

 or 

t

 button to select the first character of the 

calibration password.
When the correct first character is being displayed, press 
and release the SEL button to proceed to the next password 
character.
Repeat 1 and 2 above until the entire password is complete.
To exit the User-Defined Pass Code change mode, press and 
hold the SEL button.
Release the button when the display indicates –  –  –  –  and 
restarts in the normal operating mode.

Instructions 

F16DR, F16DRN

Pass Code Configuration

Pressure Calibration—continued

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Содержание Cecomp F16DR Series

Страница 1: ...erence Output Specify I 4 20 mA V 0 2 V BV 2 V for ranges using 15 psig sensor only psi PSI inHg INHG oz in2 ZIN inH2O INH2O ftH2O FTH2O mmHg MMHG torr TORR mmH2O MMH2O kg cm2 KGCM g cm2 GCM kPa KPA MPa MPA mbar MBAR bar BAR cmH2O CMH2O atm ATM Range see table at left F16DRN NPT 0 75 SEL TEST t s X X X X X 88888 2 Cecomp Electronics Made in USA 3 38 Standard 3 5 NEMA 4X H H H H H H H H H H H H H H...

Страница 2: ...gauge rear This cable accommodates both the gauge power supply and retransmission output Route the wires away from heat sources and moving equipment See wiring diagram Instructions F16DR F16DRN Installation Precautions RED BLACK GREEN WHITE Optional shield connection Power Power Remote Device com Types of Gauges When power is first applied the gauge proceeds through a startup sequence During the s...

Страница 3: ...ton briefly to proceed to pressure calibration or hold the SEL button for 2 seconds to save and exit The pressure calibration procedure adjusts the display indica tion and the retransmission output to correspond to the actual applied pressure If the applied pressure is not being displayed press and release the SEL button to step to the pressure calibration sequence indicated by CAL on the display ...

Страница 4: ...onds Enter Factory Pass Code 1220 Use the s and t buttons to set the left most digit to 1 Press and release the SEL button to index to the next position The 1 will remain and the second position will be blinking Use the s and t buttons to select 2 Press and release the SEL button to index to the next position 1 2 will remain and the third position will be blinking Use the s and t buttons to select...

Отзывы: