Accessing the Internal Buttons

Remove the 6 Phillips screws on the back of the unit. 
The two internal buttons are located near the lower right and left 

corners of the circuit board.

Engineering Unit Selection

Engineering unit selection (except compound ranges) is done 

via internal buttons to help prevent accidental or unauthorized 

changes. The selected engineering unit is stored in non-volatile 

memory and will be retained even with the gauge off or batteries 

removed. The available engineering units depend on the sensor 

range and display resolution.
Compound (inHg/PSIG) gauges must be changed to display single-

unit vacuum/pressure readings in the Advanced Configuration mode 

before different engineering units can be selected.
The default engineering units are mathematically converted to the 

newly selected engineering unit. When the gauge is powered up, 

the originally configured range is displayed and then the conversion 

with the selected engineering unit is displayed.
Powered the gauge up by holding the front button for 1 second.
Press and hold the internal UP button. 
Release the button when the engineering units begin to flash. 
Use the UP and DOWN buttons to scroll through the list of engineer-

ing units available for the pressure range of the sensor.
When the desired units are displayed, press and release the front 

button to save the selection and return to normal operation.
Note: The gauge will automatically revert to normal operation if no 

buttons are operated for approximately 15 seconds. 
Replace the rear cover taking care not to pinch the power wires 

between the cover and the case.

Instructions 

F16AD, F16ADN

User configuration allows requires a pass code for access and 

allows more features to be configured.
Remove the rear cover to gain access to the buttons located near 

the lower right and left corners of the circuit board.
With the gauge off, press and hold the UP button. Then press the 

front button. Release all buttons when the display indicates CFG 

and the program version then the full-scale range is indicated and 

the display is tested.
The display then indicates _ _ _ _ with the first underscore blink-

ing, and with CFGPC (configuration pass code) on the character 

segments. 
Note: The gauge will automatically revert to normal operation if 

no buttons are operated for approximately 15 seconds. To cancel 

and return to normal operation, press and release the front button 

without entering any pass code characters.

Pass Code Entry

The factory default is 3510, but this may be changed by the user 

under the Pass Code Configuration section.

1.  Use the UP or DOWN buttons to set the left-most digit to 3.
2.  Press and release the front button to move to the next position. 

The 3 will remain, and the second position will be blinking.

3.  Use the UP or DOWN buttons to select 5.
4.  Press and release the front button to index to the next position. 

35 will remain, and the third position will be blinking.

5.  Use the UP or DOWN buttons to select 1.
6.  Press and release the front button to index to the next position. 

351 will remain, and the fourth position will be blinking.

7.  Use the UP or DOWN buttons to select 0.
8.  Press and release the front button to proceed with configuration 

procedures. 

If an incorrect pass code is entered, the gauge will return to the 

start of the pass code entry sequence.

Factory/User Configuration

The upper display section will be blank, and the lower section will 

display either USER_ or FCTRY.
If USER_ is selected, the user configuration can be modified as 

described in the following steps. 
To select USER_, press and release the DOWN button. 
The lower display will indicate USER_. 
Press and release the front button to continue.
If FCTRY is selected, the existing user configuration will be replaced 

by the original factory configuration. 
To select FCTRY, press and release the UP button. 
The lower display will indicate FCTRY. 
Press and release the front button to restore the factory configura-

tion and restart the gauge.

Max/Min Configuration

Use the UP and DOWN buttons to select from the following:
MX/MN  Both highest and lowest values will be captured
MX/---  Only highest value will be captured
--/MN  Only lowest value will be captured
--/-- 

Capture feature is disabled

Press and release the front button to move to the next parameter.

Max/Min Memory

The upper display section will indicate clr.
Use the UP and DOWN buttons to select from the following:
AUTO  Automatically clear max. and min. values when the 

 

gauge is powered off

MAN 

Manually clear max. and min. values

Press and release the front button to move to the next parameter.

Gauge Type Configuration

This will only appear with 15, 100, or 200 psig ranges that were 

originally ordered as compound gauges. 
Use the UP and DOWN buttons to select from the following:
-/+EU  Vacuum is indicated as negative pressure in the 

 

selected engineering units

CMPND  Vacuum is negative INHG, pressure is PSIG. This setting 

 

will disable engineering unit selection.

Press and release the front button to save the user configuration 

and restart the gauge.
Replace the rear cover taking care not to pinch the power wires 

between the cover and the case. 

Advanced Configuration

Basic Configuration

All gauges are factory calibrated using NIST traceable calibration 
equipment. Calibration is not required before using the gauge. 
Calibration intervals depend on your quality standards, but annual 
re-calibration is customary.
Calibration equipment is not required to zero gauge reference 
ranges. Absolute reference ranges may be zeroed with application 
of full vacuum. 
Span calibration should only be performed using appropriate cali-
bration procedures with calibration standards that are at least four 
times more accurate than the gauge being calibrated.
The calibration system must be able to generate and measure pres-
sure/vacuum over the full range of the gauge. A vacuum pump able 
to produce a vacuum of 100 microns (0.1 torr or 100 millitorr) or 
lower is required for vacuum and absolute gauges.
Connect gauge to a 8-24 VAC 50/60 Hz or 9-32 VDC power supply.
Allow the gauge to equalize to normal room temperature for 
approximately 20 minutes before calibration.
Remove the rear cover to gain access to the UP and DOWN buttons 
located near the lower right and left corners of the circuit board.

Continued on next page...

Calibration Preparation

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Содержание Cecomp F16AD

Страница 1: ... reference pressure VAC gauge reference vacuum A absolute reference If vacuum gauge requires a minus sign please specify The range code indicates the gauge s default range Engineering units may be changed to any of those listed under the same sensor range The listed ranges are rounded off Range see table at left H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H H ...

Страница 2: ...isplay is tested and then oooo is displayed The gauge is now zeroed Release the button and the full scale range in the selected units are indicated the display is tested again then the actual pressure and units are displayed Attempting to zero the gauge with pressure greater than approxi mately 3 of full scale pressure or vacuum applied will result in an error condition and the display will altern...

Страница 3: ... DOWN buttons to select 0 8 Press and release the front button to proceed with configuration procedures If an incorrect pass code is entered the gauge will return to the start of the pass code entry sequence Factory User Configuration The upper display section will be blank and the lower section will display either USER_ or FCTRY If USER_ is selected the user configuration can be modified as descr...

Страница 4: ...ly full scale vacuum The character display will alternate between SPAN and CAL Adjust for a display indication of full scale vacuum using the UP and the DOWN buttons Apply 50 full scale vacuum The character display will alternate between MID and CAL Adjust for a display indication equal to 50 of full scale vacuum using the UP and the DOWN buttons Absolute Reference Gauges Apply full vacuum to the ...

Отзывы: