Precautions

WARNING! All wiring must be performed by a qualified electri-
cian or instrumentation engineer. See diagram for terminal des-
ignations and wiring examples. Consult factory for assistance. 
WARNING! Avoid shock hazards! Turn signal input, output, and 
power off before connecting or disconnecting wiring, or remov-
ing or installing module.

Précautions

ATTENTION! Tout le câblage doit être effectué par un élec-
tricien ou ingénieur en instrumentation qualifié. Voir le dia-
gramme pour désignations des bornes et des exemples de 
câblage. Consulter l'usine pour assistance.
ATTENTION! Éviter les risques de choc! Fermez le signal 
d'entrée, le signal de sortie et l'alimentation électrique avant 
de connecter ou de déconnecter le câblage, ou de retirer ou 
d'installer le module. 

Socket and Mounting

The module installation requires a protective panel or enclosure. 
Use API 008 or finger-safe API 008 FS socket.
The socket clips to a standard 35 mm DIN rail or can be 
attached to a flat surface using the two mounting holes.

Electrical Connections

Polarity must be observed for output wiring connections. If the 
output does not function, check wiring and polarity.
Each product is factory configured to your exact input and 
output ranges as indicated on the serial number label. 

Signal Input Terminals

The potentiometer must be connected to all three signal input 
terminals as shown. 0-100% of the potentiometer range must 
be used. The API 4003 G I utilizes a stable 1 VDC source to 
excite the potentiometer. 

Signal Output Terminals

Polarity must be observed when connecting the signal output to 
the load. The positive connection (+) is connected to terminal 7 
and the negative (–) is connected to terminal 8. 
Note that with a current output the module provides power to 
the output loop unless option EXTSUP was ordered for a sinking 
output requirement.

Module Power Terminals

Check  model/serial number label for module operating voltage 
to make sure it matches available power.
AC power is connected to terminals 1 and 3. 
For DC powered modules, polarity MUST be observed. 
Positive (+) is wired to terminal 1 
Negative (–) is wired to terminal 3 

Installation and Setup 

API 4003 G I

Potentiometer Input

 Terminal

Zero or low end of potentiometer

 4 

Potentiometer wiper arm

 5

Full scale or high side of potentiometer

 6 

(

+1 VDC)

API 008 FS

 

8-Pin Finger Safe Socket

1.063 

[27.00]

2.559 

[65.00]

0.126 

[3.20] 

Dia.

1.181 [30.00]

0.220 [5.60]

8

7

6

5

4

1

2

3

1.496 [38.00]

Contacts: Nickel plated 

brass, 10 Amps max.
Screws: M3

zinc 

plated steel
Wiring: #14 AWG solid 

or stranded

Top View

Terminal Location

8

7

6

5

4

1

2

3

API 008

 

8-Pin Socket

0.165 

[4.19] 

Dia.

1.06 

[26.9]

2

1

3

4

5

6

7

8

0.312 [7.92]

0.98 

[24.9]

1.295 

[32.89]

1.61 [40.9]

Contacts: Nickel plated 

brass, 10 Amps max.

Screws: 6-32 x 1/4", zinc 

plated steel

Wiring: #12 to #22 AWG 

solid or stranded, or 1/4" 

or 5/16" spade lug

2.11 

[53.6]

Terminal Location

Top View

8

7

6

5

4

1

2

3

Calibration

The API 4003 G I comes from the factory calibrated to your 
specifications. Field calibration is typically not required, how-
ever, Zero and Span potentiometers are available to fine-tune 
the module output. 
1.  Apply power to the module and allow a minimum 20 minute 

warm up time.

2.  Set the input potentiometer to its minimum value to provide 

a minimum input to the module.

3. Connect an accurate measurement device to the module

output. Adjust the module’s Zero potentiometer for the
exact minimum output desired. The Zero control should
only be adjusted when the input signal is at its minimum
to produce the corresponding minimum output signal. For
example: 4 mA for a 4-20 mA output or –10 V for a ±10V
output.

4. Set the input potentiometer at its maximum, and then

adjust the module’s Span pot for the exact maximum output 
desired. The Span control should only be adjusted when the 
input signal is at its maximum. This will produce the cor-
responding maximum output signal. Example: for 4-20 mA
output signal, the Span control will provide adjustment for
the 20 mA or high end of the signal.

5.  Repeat adjustments for maximum accuracy.

Output Test Function

When the Test button is depressed it will drive the output with a 
known good signal that can be used as a diagnostic aid during 
initial start-up or troubleshooting. When released, the output 
will return to normal.
The Test Cal. level is factory set to approximately 50% output. 
Example: If you are checking a 4-20 mA current loop, when the 
push button is held depressed, the output from the module will 
be approximately 12 mA.

Operation

The API 4003 G I utilizes a stable 1 VDC source to excite the 
potentiometer. This voltage is stabilized against the potentiom-
eter resistance value variations over the entire operating range. 
The resulting potentiometer wiper voltage is amplified and 
passed through an optical coupler to the output stage where it 
is scaled to the desired output range.
The green LoopTracker

®

 input LED provides a visual indication 

that a signal is being sensed by the input circuitry of the mod-
ule. It also indicates the input signal strength by changing in 
intensity as the process changes from minimum to maximum.
If the LED fails to illuminate, or fails to change in intensity as 
the process changes, check the module power or signal input 
wiring. Note that it may be difficult to see the LEDs under bright 
lighting conditions.
The red LoopTracker output LED provides a visual indication 
that the output signal is functioning. It becomes brighter as 
the input and the corresponding output change from minimum 
to maximum. 
For current outputs, the red LED will only light if the output loop 
current path is complete. For either current or voltage outputs, 
failure to illuminate or a failure to change in intensity as the 
process changes may indicate a problem with the module 
power or signal output wiring. 

To maintain full isolation 

avoid combining power 

supplies in common with 

output and unit power.

8

7

(–)

(+)

PLC, Display, 

Recorder w. 

Voltage Input

Voltage output

AC or DC (–) 

Module power 

AC or DC 

(+)

Socket top 

view

2

1

8

7

3

4

5

6

(–)

(+)

Sourcing mA 

output  

20 V loop power

PLC, Display, 

Recorder w. 

mA Input

R

i

Key down 

when 

panel 

mounting

Full Scale

 

Wiper Arm

 

Minimum

Potentiometer Input

Potentiometer Full Scale 

and Minimum may be 

switched for reverse 

output

– 

+

(–)

(+)

EXTSUP Option 

Current 

Sinking mA 

Output

PLC, Display, 

Recorder w. 

mA Input

R

i

8

7

Loop 

Power 

Source

8

7

(–)

(+)

Powered  

mA Device

EXTSUP Option 

Current 

Sinking mA 

Output

Externally powered device 

that provides power to the 

current loop.

External 

Power 

Source

API maintains a constant effort to upgrade and improve its products. 

Specifications are subject to change without notice. See api-usa.com for 

latest product information. Consult factory for your specific requirements.

WARNING:

 This product can expose you to chemicals includ-

ing lead and nickel, which are known to the State of California 

to cause cancer or birth defects or other reproductive harm. 

For more information go to www.P65Warnings.ca.gov

www.mod-tronic.com | [email protected] | 1-800-794-5883