background image

Wiring, Installation, Calibration 

APD 4059

M01 Option: Internal Calibration Resistor

The APD 4059 M01 has a user-specified internal calibration 
resistor. A switch on the front of the module allows switching 
of the APD's internal calibration resistor in or out of the circuit. 
The sensor manufacturer should provide the percentage of full-
scale output for the transducer when using the APD's internal 
resistor for calibration. 

M02 Option: Load Cell Calibration Resistor

The APD 4059 M02 has provisions for a load cell with its own 
calibration resistor. A switch on the front of the module allows 
switching of the load cell internal calibration resistor in or out 
of the circuit. 
Refer to the load cell manufacturer's specifications and the 
wiring diagram when connecting a transducer with its own 
internal calibration resistor. 
The transducer's calibration resistor wires are connected to 
terminals 5 and 11 on the APD 4059. 
If the transducer only has one calibration resistor wire, connect 
it to terminal 5.

Input

Refer to strain gauge manufacturer’s data sheet for wire color-
coding and identification. Polarity must be observed when 
connecting inputs.
CAUTION: Do not miswire the load cell and never short the 
excitation leads together. This will cause internal damage to 
the module.

No Sense Leads: 

If no sense leads are used, jumper terminals 

6 and 12.

With Sense Leads: 

Some bridges or load cells have one or two 

sense leads. This allow compensation for leadwire resistance. 
Connect the sense leads if used. Polarity must be observed.
Never jumper terminals 6 and 12 when using sense leads.

Output

Polarity must be observed when connecting the signal output.
If your device accepts a current input, determine if it provides 
power to the current loop or if it must be powered by the APD 
module. Use a multi-meter to check for voltage at the device's 
input terminals. Typical voltage may be 9-24 VDC.

Module Power

Check model/serial number label for module operating voltage 
to make sure it matches available power. Connect power last.
When using DC power, either polarity is acceptable, but for 
consistency with similar API products, positive (+) can be wired 
to terminal 13 and negative (–) can be wired to terminal 16. 
Connect I/O wiring before power wiring.
CAUTION: To maintain full isolation avoid wiring DC power 
supplies in common with output and unit power.

Type of Device for Output

– Term.

 + Term.

mA (current) input device that powers 

the current loop. Switch E set to “I”. 

2 (

–)

(

+)

mA (current) input device that is pas-

sive. APD module provides the loop 

power. Switch E set to “I”. 

3 (

–)

(+20 V)

Measuring/recording device accepts a 

voltage input. Switch E set to “V”. 

3 (

–)

(

+)

14

13

15

16

13  Power AC or DC 

+

14 

Earth Ground

16  Power AC or DC –

Module Power

Current sinking output 

switch E set to “I”

External device provides 

power to output loop

2

1

3

4

– 

+

+

 –

Loop 

Power 

Source

+

4-20 mA 

Device

R

i

Current sourcing output  

switch E set to “I”  

+20 V at terminal 4

2

1

3

4

Passive 

4-20 mA 

Device

R

i

– 

+20V

Voltage output  

switch E set to “V”

2

1

3

4

Voltage 

Device

– 

+

Output Wiring

M02 option 

for load cell 

with built-in 

cal. resistor. 

Connect to 

terminals 5 

and 11

No Sense Leads

Jumper 6 to 12 

ONLY if sense leads 

are NOT used

6

5

7

8

10

9

11 12

Strain 

Gauge

Span

Zero

Test

Output

Input

APD 4059

Strain Gauge to DC

Isolated Transmitter

Excitation

V

0

 

+

V

0

 –

V

EX

 –

V

EX

 

+

Exc 

+

Exc –

Sig 

Sig 

+

Test Cal.

Sig 

+

Exc –

Sig –

Exc 

+

Cal.

With Sense Leads

Sense 

+

 

Sense –

Sense 

+

 

Sense –

6

5

7

8

10

9

11 12

Strain 

Gauge

Span

Zero

Test

Output

Input

APD 4059

Strain Gauge to DC

Isolated Transmitter

Excitation

V

0

 

+

V

0

 –

V

EX

 –

V

EX

 

+

Exc 

+

Exc –

Sig 

Sig 

+

Test Cal.

Sig 

+

Exc –

Sig –

Exc 

+

Cal.

M02 option 

for load cell 

with built-in 

cal. resistor. 

Connect to 

terminals 5 

and 11

Cu 60/75°C conductors 14 AWG max.

Mounting to a DIN Rail

Install module vertically on a 35 mm DIN rail in a protective 

enclosure away from heat sources. Do not block air flow. 
Allow 1" (25 mm) above and below housing vents for air 

circulation.
1. Tilt front of module downward and 

position against DIN rail. 

2.  Clip lower mount to bottom edge of 

DIN rail.

3.  Push front of module upward until 

upper mount snaps into place.

Removal

1.  Push up on the bottom back of the 

module.

2.  Tilt front of module downward to 

release upper mount from top edge 

of DIN rail. 

3.  The module can now be removed 

from the DIN rail.

api-usa.com

1220 American Way Libertyville, IL 60048 
Phone: 

800-942-0315

 

Fax: 800-949-7502

A

A

BSOLUTE

BSOLUTE

 P

 P

ROCESS

ROCESS

 I

 I

NSTRUMENTS

NSTRUMENTS

Module powers mA output loop

Basic Calibration

The Zero, Span, and Excitation potentiometers are used to cali-

brate the output. This calibration procedure does not account 

for offsets or tare weights. If your system has an offset, tare 

weight or deadweight, refer to the Offset Switch procedure. 
Input and output ranges, if specified on your order, are factory 

pre-configured (at 24°C ±1°C). 
Note: Perform the following calibration procedure any time 

switch settings are changed.
To achieve optimum results, the system should be calibrated 

using an accurate bridge simulator, pressure calibrator, or 

calibration weights depending on the application.
1.  Apply power to the module and allow a minimum 20 minute 

system warm up time.

2.  Using an accurate voltmeter across terminals 10 and 12, 

adjust the Excitation voltage potentiometer on front of the 

APD 4059 for the exact voltage desired.

3.  With the input set at zero or the minimum, adjust the front 

Zero pot for a zero or low-end output (for example, 4 mA for 

a 4-20 mA output or –10 V with a ±10 V output). 

4. The zero pot may also be adjusted for a zero reading on 

the output display instrumentation, e.g. control system or 

process indicator. Adjusting the zero pot this way eliminates 

calibration errors in the display instrumentation.

5.  Set the input at maximum, and then adjust the Span pot for 

the exact maximum output desired. The Span control should 

only be adjusted when the input signal is at its maximum. 

Push-Pull Calibration

1.  Apply power to the module and allow a minimum 20 minute 

system warm up time.

2.  Using an accurate voltmeter across terminals 10 and 12, 

adjust the Excitation voltage potentiometer to the exact volt-

age desired.

3. Connect the precision resistor between Exc+ and Sig+. This 

will simulate the cell under tension and apply negative volt-

age to the input.

4. Adjust the Zero control to –80% output since the resistor is 

scaled for 80% of defection.

5. Remove the precision resistor.
6. Connect the precision resistor between Exc– and Sig–. This 

will simulate the cell under compression and apply a posi-

tive voltage to the input.

7. Adjust the Span control for +80% output since the resistor 

is scaled for 80% of defection.

8. Remove the precision resistor. The output should be near 

0 V. It is possible for zero to be off a small amount due to 

stacking of tolerances within both the load cell, wiring, and 

the APD 4059. 

 * Do not make connections to unused terminals!

Colors shown are an example only.  

See manufacturer's specifications for wiring designations.

Shield wires should be grounded at one end only

To maintain full isolation, avoid combining power supplies 

in common with input, output, or unit power.

1

 2 3 4

5  6 

7  8

9 10 11 12

13 14 

15 

16

Input

Cal. and Sense

Output

Power

Wire terminal 
torque 
0.5 to 0.6 Nm or 
4.4 to 5.3 in-lbs

 * Do not make 

connections to 

unused terminals!

 * Do not make connections 

to unused terminals!

1 strain gauge shown. Connect up to 4 in parallel if all leads are 

equal length. Unequal length leads or strain gauges with calibra-

tion variances may require sum box SG-EQ4 to aid in equalization.

Spring 

Clip

Upper 

Mount

Lower 

Mount

Содержание APD 4059 Series

Страница 1: ...ors IP 40 requires installation in panel or enclosure For use in Pollution Degree 2 Environment Mount vertically to a 35 mm DIN rail Four 4 terminal removable connectors 14 AWG max wire size Dimensions 0 89 W x 4 62 H x 4 81 D 22 5 mm W x 117 mm H x 122 mm D Height includes connectors Power Standard 85 265 VAC 50 60 Hz or 60 300 VDC D option 9 30 VDC either polarity or 10 32 VAC Power 2 to 5 Watts...

Страница 2: ...only purpose is to adjust or cancel effects of the low end of the input range not corresponding nominally to 0 mV Setting this switch to 0 results in no offset 2 To RAISE the output zero rotate switch C from 1 thru 7 until the Zero control can be set for your application 3 To LOWER the output zero rotate switch C from 9 thru F until the Zero control can be set for your application 4 If switch posi...

Страница 3: ... 11 Cu 60 75 C conductors 14 AWG max Mounting to a DIN Rail Install module vertically on a 35 mm DIN rail in a protective enclosure away from heat sources Do not block air flow Allow 1 25 mm above and below housing vents for air circulation 1 Tilt front of module downward and position against DIN rail 2 Clip lower mount to bottom edge of DIN rail 3 Push front of module upward until upper mount sna...

Страница 4: ... transducer must be con nected per the manufacturer s specifications The sensor manufacturer should provide the percentage of full scale output for the transducer when using a calibration resistor This is often 80 of maximum output 1 Apply power to the module and allow a minimum 20 minute system warm up time 2 Using an accurate voltmeter across terminals 10 and 12 adjust the Excitation voltage pot...

Отзывы: