MOISTURE DURING THE SUMMER MONTHS 

The amount of moisture, condensation, or high humidity related issues increase during the summer 
and, in most cases, will self-resolve when the weather cools down.  Please note a refrigeration 
system will NOT generate moisture or water but simply condenses the moisture that is already in 
the chamber. Keeping the unit in an air conditioned, low humidity space will resolve many issues. 
Other things you should check: 

1.

 

Location of the refrigerator (See Quick Troubleshooting Guide above) 

2.

 

Door sealing and frequency of door opening event (See Quick Troubleshooting Guide above) 

3.

 

Make sure there is a water trap (U-shaped loop) in the drain tube near the end. This will 
“trap” a small amount of water in the loop and prevent air from entering the chamber 
through the tube.  
 

BEFORE CALLING THE MANUFACTURER’S TECHNICAL SUPPORT DEPARTMENT, 

please have the 

unit’s model and serial number ready as well as the problem description.  The model and serial 
number is located on the serial tag which can be found on the interior left upper wall of the 
unit. 

PROBLEM: 

POSSIBLE CAUSE / SOULTIONS: 

Compressor noises 

 

Compressor may be overheated.  Please check the room 
temp and ensure the range is within 65°F to 85°F.  If the 
problem still exists, call for service. 

Moisture collects inside 

 

Door gasket is not sealing properly. Check for debris, 
cracks, and items passing through door at the gasket.                                                             

 

The refrigerator or freezer is facing a doorway or is 
underneath of air conditioning vent. Relocate the unit or 
redirect air vent. 

 

Too many door openings. Minimize time door is open. 

 

Hot, humid weather increases condensation. 

 

Make sure there is a water trap (U-shaped loop) in the 
drain tube near the compressor. This will “trap” a small 
amount of water in the loop and prevent air from 
entering the chamber through the tube.  

Moisture collects on outside 
surface 

 

Hot, humid weather increases condensation. 

 

As humidity decreases, moisture will disappear. 

Odor inside the unit 

 

Interior needs to be cleaned.  See section on 
maintenance and cleaning in this manual.  

 

Make sure product containers are tightly sealed to 
prevent leakage 

Door will not close 

 

The unit is not level. Refer to the Leveling section at the 
beginning of this manual 

 

Check for dirt and debris or items passing through the 
door seal. 

Содержание PH-ABT-UCFS-0220M

Страница 1: ...er at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is being unpa...

Страница 2: ...have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left upper wall of the unit...

Страница 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Страница 4: ...eptable and is recommended on this unit POWER SOURCE The supply circuit to this cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle...

Страница 5: ...as the excess heat has been removed TEMPERATURE CONTROL NOTE The control is preset at the factory to provide the desired air and product temperatures inside of the unit and requires no further adjust...

Страница 6: ...Leave about 2 3 inches of free space along the back and sides of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocking the evap...

Страница 7: ...properly Amount of stored product is overloaded Product replacements are pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the produ...

Страница 8: ...the interior left upper wall of the unit PROBLEM POSSIBLE CAUSE SOULTIONS Compressor noises Compressor may be overheated Please check the room temp and ensure the range is within 65 F to 85 F If the p...

Страница 9: ...and their seating surfaces with a clean soft cloth Shelves Soap and water Do not wash removable shelves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not use commerc...

Страница 10: ...e paid by the customer ADDITIONAL THREE YEAR COMPRESSOR WARRANTY In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compr...

Страница 11: ...t damage during transit or installation fire flood or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with t...

Отзывы: