manualshive.com logo in svg
background image

  

 

P a g e

 | 10 

At each setting, press [UP] or [DOWN] to make a change.  Then press [MENU] to 
confirm the new change, and move to the next setting. 
 
For example, to change the low temp alarm point: press [MENU] till the display 
shows [ALL].  Press [UP] to change the factory setting 0C to 1C.  Press [MENU] to 
confirm the setting.  And the display will show “Ad” next. 
 
To  return  to  the  main  display,  either  cycle  through  the  entire  menu  from  “ALU”  to 
“OFF”,  or  simply  take  no  action  for  30  seconds.    The  display  will  return  to  the 
temperature display mode. 
 
HIGH TEMP ALARM POINT (ALU) 
Under this function, you are able to change the high temperature alarm point.  Once 
the refrigerator temperature reaches the set point, D-100 will display “HA” message, 
and send out the audible alarm. 
 
LOW TEMP ALARM POINT (ALL) 
Under this function, you are able to change the low temperature alarm point.  Once 
the refrigerator temperature reaches the set point, D-100 will display “LA” message, 
and send out the audible alarm. 
 
VISUAL & AUDIBLE ALARM DELAY (Ad) 
This function allows users to adjust the time delay between the occurrence of an error 
and the visual & audible alarm’s activation.  For example, if the setting is 1 (minute), 
when the temp warms above the high temp alarm point, it would take 1 minute before 
the visual and audible alarm are activated. 
 
We recommend to keep the setting at 0 (minute). 
 
CHANGE TEMPERATURE READOUT BETWEEN °C AND °F

 

Press [MENU] 4 times till “CF” displays.  Press [UP] or [DOWN] to change between °C 
and °F.  The right side °C / °F indicator will toggle according to your selection.  Press 
[MENU] again to confirm your change. 
 
DOOR AJAR ALARM (dA) 
This feature is disable on most models (Setting: NO).  Only selected models have the 
door alarm switch installed, and feature unlocked.  If your refrigerator model does not 
have the door alarm switch installed, and dA setting is turned on to 0 to 5 (minutes), 
the display will give out dA (door ajar) alarm even when the door(s) is closed.   
 
If your refrigerator has the door alarm switch(s) installed, you can change the door ajar 
alarm  duration  from  0  minute  to  5 minutes.  If  one  or more  doors  is  opened  out  of 
duration time period, the display will activate “dA” message.  Once the door is closed, 
the alarm will be turned off. 

Содержание ABT-FRP-04

Страница 1: ...or Temp Control V5 2 Table of Contents RECEIVING AND SHIPPING DAMAGE HANDLING 2 SAFETY 3 RELEASE OF LIABILITY 3 ELECTRICAL INFORMATION 4 INSTALLATION 5 OPERATION 7 Quick Troubleshooting Guide 16 Maint...

Страница 2: ...eir contact information can be found on the carrier s web site The second type of shipping damage is concealed damage This type of damage will probably not be apparent at time of delivery and may not...

Страница 3: ...reezer This is a flammable material storage refrigerator or freezer All laboratory safety rules must be followed during the operation of this refrigerator or freezer RELEASE OF LIABILITY Before you st...

Страница 4: ...to NEC National Electrical Code Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection SUPPLY VOLTAGE SHOULD NOT VARY MORE THAN 5 FROM SERIAL PLAT...

Страница 5: ...ntific medical application Many components are heavy duty and extra sized in order to meet the ultimate temperature performance Therefore the sounds generated from its operation may not be accepted by...

Страница 6: ...IELD INSTALLATION Each refrigerator or freezer is equipped with a 3 8 probe access port hole for your independent probe installation The port hole is generally located in the back of your refrigerator...

Страница 7: ...p alarm function D 100 temp display and alarm system is specifically designed to display the refrigerator s interior sample temperature and give out alarms if an error occurs FEATURES Sample temperatu...

Страница 8: ...tteries will support the alarm system and the temperature display for 8 hours When the power resumes the batteries will be automatically recharged During a power failure event the visual display will...

Страница 9: ...point MENU Press MENU button to enter the user adjustable settings Press UP or DOWN to select which setting you would like to make an adjustment CODE DESCRIPTION FACTORY DEFAULT ALU High temp alarm p...

Страница 10: ...unction allows users to adjust the time delay between the occurrence of an error and the visual audible alarm s activation For example if the setting is 1 minute when the temp warms above the high tem...

Страница 11: ...ome time to correct an error before sending the alarm to the remote monitor system For example rD is set to be 20 minutes A user forgets to close the refrigerator s door and the temperature warms up t...

Страница 12: ...d allows the user to easily select the display units set point and differential set point Please Note The digital temperature controller has been factory set and tested to allow your unit to operate a...

Страница 13: ...urn on and off in order to maintain the cold temperature in the storage chamber In this controller the point where the compressor is cut off is called SET POINT The point where the compressor is turne...

Страница 14: ...t that SET or ALL could reach 1 C or 34 F refrigerator 30 C or 22 F freezer Change the set point compressor turn off point Press and hold SET until C or F icon blinking Press UP or DOWN to change the...

Страница 15: ...e until Hy appears on the display Press UP or DOWN until Lod shows up Press SET to enter the setting Press UP or DOWN to toggle between the air temp P1 or the sample temp P3 Press SET once to confirm...

Страница 16: ...refrigerator or freezer may not be able to maintain the interior temp at proper range Door is not closed properly Amount of stored product is overloaded Product replacements are pushed against rear wa...

Страница 17: ...ther increases condensation Make sure there is a water trap U shaped loop in the drain tube near the compressor This will trap a small amount of water in the loop and prevent air from entering the cha...

Страница 18: ...3 Make sure there is a water trap U shaped loop in the drain tube near the end This will trap a small amount of water in the loop and prevent air from entering the chamber through the tube BEFORE CALL...

Страница 19: ...helves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not use commercial household cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smooth handle...

Страница 20: ...ondenser clean you will minimize your service expense and lower your electrical costs The condenser requires scheduled cleaning every 30 day or more frequently in dirty environments Proper cleaning in...

Страница 21: ...re to replace the louvered grill as it provides protection for the condenser 8 Reconnect the electrical power to the unit Hydrocarbon Service Notes According to U S Code of Federal Regulation 40 Part...

Страница 22: ...r repaired parts will be warranted for only the unexpired portion of the original warranty Horizon Scientific Inc will not assume any shipping or cartage costs for parts under warranty These costs sha...

Страница 23: ...conformities from any cause Purchaser further acknowledges that the purchase price of the Product reflects these warranty terms and remedies ALTERATION NEGLECT ABUSE MISUSE ACCIDENT DAMAGE DURING TRA...

Страница 24: ...er at 800 648 4041 Horizon Scientific Inc s shipping records showing date of shipment shall be conclusive in establishing the warranty period All claims should include model number of the refrigerator...