Errors
33
7.
Errors
If an error occurs during the communication between Hydrotracer and PC, the HydroTracer
sends an error message to the software. In some cases, this will interrupt the measurement.
Following table shows the possible errors:
Code
Error
2
Reactor Heater
4
Door Contact Reactor
16
PT1000 Mainboard (PTC3)
32
BME280 – Temperature
64
BME280 – Humidity
128
BME28ß – Pressure
256
Communication Mainboard <-> PC
512
Communication Mainboard <-> Sensorboard
1024
general Sensorboard
If the connection between PC and HydroTracer is interrupted for more than 4 minutes, the
measurement is automatically cancelled and the HydroTracer cools down.