D.
Next add in which receiver coils are being used
on each input channel.
For a 20x20m loop, only use the RC-5.
For a 40x40m or 100x100m transmitter loop,
the RC-5 & RC-200 can both be used.
For a 200x200m loop, only use the RC-200.
Offsets can be configured here as well as the
orientation of the RX coil. A
“
Z
”
component
measurement would be from a regular,
horizontally laid, receiver.
“
X
”
and
“
Y
”
are
when the loop is mounted vertically with
respect to the transmitter loop.
“
X
”
would
normally be with the RC antenna aligned up
the centre of the transmitter loop;
“
Y
”
would
be with it aligned across the centre of the
transmitter loop.
E. Now that the hardware is configured, the
final step is to choose a suitable script. A
table is provided below to assist with the
choice.
The
“
Number of Cycles
”
box is how many
times the script (which in itself contains
hundreds of stacked measurements
–
see
table) will be repeated. Increasing this will
increase the signal to noise ratio and help
with the deeper signals. But it lengthens the
survey time; 3-5 is acceptable under normal
conditions.
On the right-hand side, an indication of how
long the measurement will take is given.