10
CALIBRATION
3 Press
[ZERO]
, the display will now say
“CAL”
and then a 5 digit number which is a
special internal reference number. The last number(s) on the right will change up or
down slowly. When these numbers are hardly changing and/or the stable indicator is
showing press
[UNIT]
. The display will now flash
“- - - - -”
for a second and then flash
with the weight you entered.
• Place the item(s) to be used for calibration on the platform, wait for a moment and
then press [
UNIT]
. Another 5 digit number will appear on the display which you DO
NOT need to worry about.
• Press
[ON/OFF]
to turn scale off to complete and save the calibration.
If a mistake has been made, pressing
[ON/OFF]
before the end of step 3, will turn the
scale off and calibration will not have taken place.
To test the calibration was successful, turn on the scale by pressing
[ON/OFF]
, once the
scale has finished self testing and the display shows
0
, put item(s) used for calibration on
the platform and check the scale is weighing correctly. If not, repeat steps 2 & 3 again.