Technical Data
Section 3
3-28
Abbott VP 2000™ Processor Service Manual
30-608308//R1—September 2008
Water Bath Data
Drying Station
Table 3.6:
Drying Station
Performance Data
Table 3.7:
Performance Data
Table 3.5:
Water Bath Conditions
Description
Specification
Input Line Pressure
10 to 100 psi
Water Type
Deionized
Flow Thru Rate
Factory set via regulator
approximately 1 liter/min
Recirculation Rate
Approximately 3.5 liters/min
Water Bath Capacity
Approximately 1,750 ml
Water In Line
3/8 inch reinforced flexible tubing
to fit standard laboratory water tap.
Adapter required for other taps.
Drain Line
1/2 inch flexible tubing gravity feed.
The entire line
must be below
the
water level in the unit.
Description
Specification
Temperature
Adjustable 25
o
C to 80
o
C
Description
Specification
Slide Capacity
50
Unheated Reagent Vessels
12
Heated Reagent Vessels
3
Water Bath
1
Drying Station
1
Reagent Basin Protocol Volume to Cover Slide
600 ml
Reagent Basin Protocol Volume to Cover Sample
470 ml
Heated Reagent Temperature Adjustment
Basin 1: Ambient to 80
o
C
Basin 2: Ambient to 80
o
C
Basin 3: Ambient to 40
o
C
Содержание VP 2000 Processor
Страница 2: ...Introduction ii Abbott VP 2000 Procesor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 13: ...Abbott VP 2000 Processor Service Manual 1 1 30 608308 R1 September 2008 SECTION 1 ABOUT THIS MANUAL ...
Страница 14: ...About This Manual Section 1 1 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 22: ...About This Manual Section 1 1 10 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 23: ...Abbott VP 2000 Processor Service Manual 2 1 30 608308 R1 September 2008 SECTION 2 SAFETY Section 2 Safety ...
Страница 24: ...Safety Section 2 2 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 38: ...Technical Data Section 3 3 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 70: ...Installation Section 4 4 2 Abbott VP 2000 Proceessor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 97: ...Section 5 Purpose of This Section Abbott VP 2000 Processor Service Manual 5 7 30 608308 R1 September 2008 ...
Страница 144: ...Troubleshooting Section 6 6 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 146: ...Troubleshooting Section 6 6 4 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES TOC ...
Страница 270: ...Shutdown Storage and Shipping Section 8 8 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 280: ...Spare Parts and Accessories Section 9 9 2 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...
Страница 288: ...Spare Parts and Accessories Section 9 9 10 Abbott VP 2000 Processor Service Manual 30 608308 R1 September 2008 NOTES ...