Keep the left hand on the Stabilizer to maintain the Clip Applier at the angle of the tissue tract.
•
Place the right index finger on the distal post, the middle two fingers in the loop, and the right thumb on the Thumb Advancer.
•
Gently retract the Clip Applier until the Locator Wings are against the arterial wall, and slight resistance is felt.
•
While maintaining the Locator Wing position against the arterial wall, push the Thumb Advancer forward with the pad of the right
thumb until the number
“3”
can be seen in the number window.
•
A
“CLICK”
should be heard to confirm completion. Check that the number
“3”
is completely visible in the number window.
“CLICK” 3
- Sheath Split
StarClose SE
Vascular Closure System
8