17
How to Recall the LOT Number
1
To recall the glucose LOT number, Press and Hold the
button. While holding the button, Press and Hold
the button.
The Display Check shows on the
display window, then the glucose
LOT number shows with the
time, month, and day of the most
recent glucose calibration.
2
After the glucose LOT number is displayed, you can
recall the β-Ketone LOT number. Press and Release the
button or the button.
The β-Ketone LOT number
shows with the word “CODE”,
the time, month and day of
the most recent β-Ketone calibration.
▼
Note
: If the LOT number that shows on your meter’s
display window is not correct:
What It Means:
Your meter may not be calibrated for the box of test strips
you are using.
What To Do:
•
Re-calibrate your meter.
•
Check that you are using the calibrator that came in
the box of test strips you are using.
Remember, improper calibration will cause
incorrect results.
3
Press and Hold the button to
turn your meter off.
CHAPTER 4
Monitoring Your
Blood Glucose
What You Will Need
•
Blood glucose test strip with its instructions for use.
•
Optium Xceed meter calibrated to match the LOT number
of the blood glucose test strip you are using.
•
Lancing device and a new, sterile lancet.
Important Information about Monitoring Your
Blood Glucose
•
Do not
use out-of-date test strips. Check the expiry date
printed on the test strip box and on each test strip foil
packet. If only the month and year
are printed on the test strip, then
the expiry date is the last day of
that month.
•
For more detailed information
about your test strip, please read
its instructions for use before
monitoring.
•
Use the test strip
immediately
when you take it out of its
foil packet.
•
Do not
use a wet, bent, scratched, or damaged test strip.
•
Do not
use the test strip if its foil packet has a puncture or
tear in it.
•
Before you monitor, allow your meter and test strip to
reach the recommended operating range of the test
strip. The test strip operating range is in the “What else do
I need to know?” section of your blood glucose test strip
instructions for use.
•
Read the lancing device instructions for use.
▼
▼
L O T 13758
EXP
2010/03
C5
Example:
Expiry date
March 31, 2010
Monit
oring
Blo
od
Gluc
ose
18
Questions? Call
Customer Support:
UK: 0500 467 466
Ireland: 1800 776633
Calibr
ating
Your
Met
er
ART13613-001_Rev-D-Guide.indd 20-21