ID NOW Instrument User Manual 59
12.2 Warnings
[
Note: These are arranged in order of the time they may appear in the process]
Example Screen:
Warning
OK
Memory Is Full
12.2.1 Home Warnings
Warning: Memory Almost Full
The instrument can run only 50 more patient tests or 9 more QC tests before the
memory is full.
Action: Press OK to continue to run the current test. With each additional test or
QC test carried out, a similar error will be displayed showing the number of tests
remaining. The Admin must export the stored test results before the memory is
full, or results will be overwritten. See Section 9.4, Export Results.
Warning: Memory Is Full
The instrument can store up to 999 patient test and 99 QC results, and is now full.
Action: To avoid overwriting stored data, the Admin must export the data. See Section
9.4, Export Results.
Alternatively, Press OK to continue to run the current test.
Note: that this will overwrite stored data.
Warning: Heater Not At Temperature
The instrument has not yet reached the required temperature.
Action: Press OK to return to the Home screen and wait for the white light on
the front of the instrument to stop flashing before trying again.
If the error persists, power off the instrument, disconnect the DC power input
connector, wait 10 seconds, reconnect and turn on the instrument. When the
white light on the front of the instrument stops flashing, try again.
If this error persists, tests cannot be run on the instrument. However, the stored
data can still be viewed:
Home>Review Memory.