System Controller |
13
BATTERY CAPACITY LIGHT
4 green lights = Approximately 75%–100% of power remains
3 green lights = Approximately 50%–75% of power remains
2 green lights = Approximately 25%–50% of power remains
1 green light = Less than 25% of power remains
Yellow diamond only = Less than 15 minutes of power remains
PUMP RUNNING LIGHT
Stays lit green as long as the LVAD is running.
HAZARD ALARM
When the red heart symbol comes on, follow the on-screen
instructions. Do this immediately.
YELLOW WRENCH
Lights up yellow when the System Controller detects a
mechanical, electrical or software issue with the system.
LOW BATTERY ALERT
Less than 5 minutes of battery power remain. Immediately
replace used batteries with fully charged batteries or switch
to the MPU.