4-1
Questions? Refer to Page 11-22 For Technical Support Instructions.
Control Test
This chapter describes the steps for running a control test with control solutions. Control tests verify the performance of the
FreeStyle Precision Pro meter and test strips.
The quality control features of the meter can be customised to fit the requirements of each facility.
The meter can be set to require:
• Tests of low, mid, and high control solutions or a variation depending on the facility’s policy.
• Tests of control solutions at relative times (e.g. every 24 hours) or fixed times (e.g. at 6:00, 14:00, and 22:00).
• Tests of control solutions after strip port module replacement.
• Control test results to appear as numeric value or Pass/Fail.
These quality control features can be set through the data management system.
Operating Guidelines for Control Solution Testing
Observe the following guidelines to obtain optimal quality control results using the FreeStyle Precision Pro
system:
• Use only MediSense Glucose and Ketone Control Solutions with compatible test strips to verify the performance of the
FreeStyle Precision Pro meter and test strips.
• When opening a new bottle, write the discard date on the bottle label. Each bottle of control solution is stable for 90 days
after opening or until the expiry date printed on the label, whichever comes first.
Chapter 4: Control Test
Содержание FreeStyle Precision Pro
Страница 1: ...Operator s Manual for Healthcare Professionals...
Страница 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Страница 98: ...7 Linearity Test...
Страница 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Страница 172: ...12 Maintenance Cleaning the Exterior Surface Replacing the Strip Port Module Replacing the Batteries...
Страница 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...