How do I know if the control solution test results are accurate?
• The control solution test results should fall within the range printed
on the test strip vial label.
Important:
The control solution range is a target range for control
solution only. It is not a target range for your blood glucose level.
• If control solution test results are out of this range,
repeat the test
.
• Out of range test results may be caused by:
– Expired or bad control solution.
– Expired or bad test strip.
– Error in performing test.
– Watered-down control solution.
– Code on test strip vial does not match code set in meter.
– Meter malfunction.
– Control solution test done outside 59º to 104ºF (15º to 40ºC).
If your control solution test results continue to fall outside the
range printed on the test strip vial:
1. The FreeStyle Flash
™
System may not be working properly.
2. Do not use the FreeStyle Flash System to test your blood glucose.
3. Call Customer Care.
Where can I obtain FreeStyle Control Solution?
• Contact Customer Care to find out where you can
obtain FreeStyle Control Solution.