9
MANUAL PRIMING
1. Gently push the lilac
prime tab on the giving set
2. Steadily apply pressure
until feed begins to flow (no
need to use a pumping action)
3. Completely fill the set by
allowing the feed to expel
air from the tubing
4. Open the pump door
5. Grasp the purple
cassette body
6. Loop pump insert around
wheel, stretching lightly
7. Gently pull down on
the loop
8. Seat lilac cassette into
the pump
9. Gently press the tubing
to ensure it is fully seated in
the slots below the cassette
10. Close the pump door. You will hear
a click when it has been closed properly
In order to prime the giving set, ensure that the end of the giving set
is free from any attachment/cap