7 SETTING UP THE SYSTEM
This section describes how to unpack the Console, connect the power cord, the Motor,
and the Flow Probe and how to mount, power up and operate the System including using
the optional elements listed in
Table 2
. The Console carries the following Factory
and
Power-Up
defaults for each specified operating parameter:
Table 8: Console Factory & Power-Up Default Values
OPERATING PARAMETER
FACTORY
DEFAULT
POWER-UP
DEFAULT
Minimum Flow Alert
0.0 LPM
0.0 LPM
Maximum Flow Alert
8.0 LPM
8.0 LPM
Flow Display range
-2.0 – 10.0 LPM
-2.0 – 10.0 LPM
Pressure Subsystem
Inactive
Last state
Pressure limits
-30mmHg / +200mmHg -30mmHg / +200mmHg
Language
English
Last state
RPM Increment
100
Last state
Flow Limit Sensitivity
Normal
Last state
Recorder Speed
2 minutes
2 minutes
7.1 Unpacking
1. Remove the Console, AC Power Cord, and Flow Probe from the Console’s shipping
container.
2. Remove the Motor from its shipping box.
3. Remove the Monitor from its shipping box
4
Factory Default: pre-selected operating parameter of a 2
nd
Generation CentriMag Console as it is shipped
from the manufacturer.
5
Power-Up Default: 2
nd
Generation CentriMag Console operating parameter after the 1
st
use. These values
are stored in the Console’s permanent memory and recalled each time the Console is powered up.
6
Note: If the probe detects forward flow of more than 10 LPM then it will display as “^^.^^ LPM”.
7
Last state: the value/state carried from the last operational use of the 2
nd
Generation CentriMag Console.