3200 System Operator’s Manual
9-45
9140181H—October 2001
Section 9
Service and Maintenance
Enabling the Analyzer
When all syringes have been re-installed, do the following:
1. Press
[ENABLE ANALYZER]
followed by
[RETURN]
,
[MAIN]
and
[RUN]
to return to the
RUN
screen.
2. Run several background counts in the Open Mode and observe the action of
the syringes during the cycle. The plungers should move smoothly up and
down and should not leak.
3. After the operation of all the syringes has been verified, do the following:
a. On the CS model, reattach the Front Skirt using the four Phillips-head
screws. Do not pinch the tubing at the bottom.
b. On the SL model, align the removable right side skirt panel over its
holding pins and slide down. If the Sample Loader has been pulled away
as described in
perform the following steps:
1. Carefully lift and push the Sample Loader toward the instrument.
2. Push the Sample loader down to lock it in place.
3. Reinstall the Phillips-head screw that holds the Sample loader Tower
in place on the Front Panel. Do not pinch any tubing between the
Tower and the Flow Panel. Refer to Figure 9.14.
4. Tighten the flat head screws: two on each side of the instrument.
Refer to Figure 9.15.
c. Reattach the Tower Cover (Tower Door on the CS model must be open)
and press into place.
d. Close the Left and Right Front Covers.
4. Run at least three background counts. Check that the background counts are
acceptable before running controls or patient samples. If the counts are
unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
5. Record this maintenance in your maintenance log.
HGB Flow Cell Manual Cleaning Procedure
NOTE:
This is not a routine cleaning/maintenance procedure. It should only be
performed if routine methods fail or at the request of the Abbott
Hematology Customer Support Center.
Under normal circumstances, the Auto-Clean cycle is sufficient to ensure the
cleanliness of the HGB Flow Cell. If the Auto-Clean cycle fails to adequately clean
the flow cell, the procedure described below may be used.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...