Nonscheduled Maintenance Procedures
Section 9
Materials Required
1. A large container filled with approximately 500 mL of warm, deionized
water
2. DYN-A-WIPE lint-free pads (or other lint-free pads)
3. Distilled water
4. Small container of each reagent to refill the clean syringes
5. Phillips-head screwdriver (#2)
6. Flat-head screwdriver
Disabling the Analyzer
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed by
>
DIS/ENAB
ANALYZER
@
and
[DISABLE ANALYZER]
2. To gain access to the syringe assemblies, open Left and Right Front Covers
as follows. Gently pull on cover using hand hold and swing door open about
30-45 degrees. Lift door fully upward from bottom area closest to door hinge
bracket, then rotate door fully open so lift and hold feature on upper door
hinge mechanism can be engaged. Lower door gently. Gently pull the Tower
Cover and lift the cover up and off the instrument. (Tower Door on the CS
model must be open.)
a. On the CS model, remove the four Phillips-head screws holding the Front
Skirt to the instrument and remove the skirt.
b. The skirt on the SL model has two removable sections adjacent to the
Flow Panel. Remove the section on the right side of the Flow Panel by
lifting it up and off its holding pins.
For easy access to the Flow Panel, the Sample Loader on the SL Model can
be pulled away from the analyzer by performing the following optional steps:
NOTE:
Instruments with serial number 59965AF and below do not have
the hardware configuration to support this option.
1. Remove the Phillips-head screw that holds the Sample Loader Tower
in place on the Front Panel. Refer to the following Figure 9.14.
Figure 9.14 Sample Loader Tower
2. Loosen the flat-head screws, two on each side of the instrument.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...