CELL-DYN
®
3200 System Operator’s Manual
9-25
9140181K—July 2002
Section 9
Service and Maintenance
Diluent/Sheath Filter Replacement Procedure
The Diluent/Sheath Filter is located on the front panel to the left of waste chamber
#3. Replace the filter once a month, or whenever contamination is suspected. A
sign of contamination is usually manifested by high platelet background, poorly
defined PLT-RBC (0°/10°) scatterplot, excessive WBC flagging, or erroneous
5-part differential results. Refer to the Troubleshooting Guide for more details.
Materials Required
1. Diluent/Sheath Filter List Number 06H92-01
2. Paper Towel
3. Hemostat
4. Gloves, Safety Glasses, Lab Coat
Procedure
1. If the instrument is not in the STANDBY mode, then press
[SPECIAL
PROTOCOLS]
, followed by
[DAILY SHUTDOWN]
. If the instrument is already
in STANDBY mode, then proceed to
Step
2.
2. Turn the power switch OFF.
3. Locate the Diluent/Sheath Filter on the front panel. Refer to Figure 9.9.
Figure 9.9 Replacing the Diluent/Sheath Filter
4. Remove lower left section cover for easier access.
5. Remove the Diluent/Sheath Filter from its spring clip.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...