CELL-DYN
®
3200 System Operator’s Manual
6-3
9140181H—October 2001
Section 6
Calibration Procedures
General Information
The CELL-DYN 3200 System has two modes of operation:
• Open Mode
• Closed Mode
Both the Open and Closed Modes are calibrated individually. It is recommended
that the primary mode (open mode) be calibrated first.
NOTE:
The procedure for calibrating Closed Mode differs slightly between the
SL and CS models.
Three methods of calibration are available on the CELL-DYN 3200 System:
Auto-Calibration, Enter Factor, and Latex.
1. The Auto-Calibration method is designed to quickly and easily calibrate the
system, using commercial control material or fresh whole blood.
a. On the CS model, Auto-Calibration can be used in both the Open and
Closed Modes.
b. On the SL model, Auto-Calibration can be used only in the Open Mode.
The Enter Factor Method must be used to calibrate the Closed Mode.
2. The Enter Factor method is a manual procedure for calibrating the
instrument. With this method, Fresh Whole Blood is normally used but
Calibrator may also be used.
3. Latex material may also be used to calibrate the instrument but this procedure
should only be performed by Abbott service personnel.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...