3-24
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Principles of Operation
Flow Cytometry
Section 3
Laboratory Worksheet Screen
The
LABORATORY WORKSHEET
screen is provided to assist the laboratory
staff in data review and validation (refer to Figure 3.11.) This screen is for
laboratory use only. This screen displays the 5-Part Differential plus additional
parameters. The
RUN
and
DATA LOG
screens display only the 5-Part Differential.
The difference between the two formats is shown in Table 3.1 and 3.2.
NOTE:
The parameters MON and LYM have an “e” after the label, indicating that
the values are estimated. MONe represents monos minus blasts. LYMe
represents reported lymphs minus variant lymphs.
Figure 3.11 Laboratory Worksheet Screen
To access the
LABORATORY
WORKSHEET
screen, press the Page Down key on the
keyboard while the
RUN
menu is displayed or while the
DISPLAY SPECIMEN
screen
(in the
DATA LOG
menu) is displayed. The format is fixed and cannot be changed.
The specimen displayed on the screen is identified by the Specimen ID number and
Sequence number in the upper left corner of the screen.
To print the
LABORATORY WORKSHEET
screen, press the [
] soft key or the
Print Screen key on the keyboard. Press [
RETURN
] to return to the
RUN
menu or
DISPLAY SPECIMEN
screen (in
DATA LOG
menu), depending on which menu was
used to access the worksheet.
The 5-Part Differential separates WBC into 5 components: Neutrophils,
Lymphocytes, Monocytes, Eosinophils, and Basophils. The additional parameters
further separate the Neutrophils, Lymphocytes, and Monocytes into their
constituent components. Eosinophils and Basophils are the same in both tables.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...